深度学习自适应波束合成提升三维成像横向分辨率
研究提出将深度学习自适应波束合成(ABLE)与大孔径阵列生成训练目标结合,并扩展感受野至多体素,用于改善微型超声探头经微波束合成后的横向分辨率。在仿真数据上训练网络即可实现,定量与定性均优于现有延迟叠加、相干因子、滤波延迟乘加和特征基最小方差波束合成器。
要点
- ABLE结合大孔径阵列训练目标可补偿微波束合成导致的聚焦损失。
- 扩展感受野至多体素进一步改善横向分辨率。
- 仅需仿真数据训练,便于实际部署。
核心结论
- 所提方法在定量和定性上均改善横向分辨率,图像质量优于现有延迟叠加、相干因子、滤波延迟乘加和特征基最小方差波束合成器。
研究设计与样本
- 研究为方法学开发与仿真验证,训练目标由大孔径阵列生成,网络仅用仿真数据训练,未报告样本量、中心数及具体设备参数。
关键影像表现/征象
- 微波束合成降低数据率和线缆数但引入聚焦损失,现有波束合成器改善不足或计算成本过高。
临床应用与阅片要点
- 在微型探头三维成像中,可考虑用ABLE替代现有波束合成器以提升横向分辨率。
- 训练目标来自大孔径阵列,部署时仅需仿真数据,便于临床转化。
局限性
- 研究基于仿真数据验证,未报告临床或体模实验数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
There is an increased desire for miniature ultrasound probes with small apertures to provide volumetric images at high frame rates for in-body applications. Satisfying these increased requirements makes simultaneous achievement of a good lateral resolution a challenge. As micro-beamforming is often employed to reduce data rate and cable count to acceptable levels, receive processing methods that try to improve spatial resolution will have to compensate the introduced reduction in focusing. Existing beamformers do not realize sufficient improvement and/or have a computational cost that prohibits their use. Here we propose the use of adaptive beamforming by deep learning (ABLE) in combination with training targets generated by a large aperture array, which inherently has better lateral resolution. In addition, we modify ABLE to extend its receptive field across multiple voxels. We illustrate that this method improves lateral resolution both quantitatively and qualitatively, such that image quality is improved compared with that achieved by existing delay-and-sum, coherence factor, filtered-delay-multiplication-and-sum and Eigen-based minimum variance beamformers. We found that only in silica data are required to train the network, making the method easily implementable in practice.
引用
- Ossenkoppele, Luijten, Bera, et al. Improving Lateral Resolution in 3-D Imaging With Micro-beamforming Through Adaptive Beamforming by Deep Learning. Ultrasound Med Biol. 2022. PMID: 36253231. doi:10.1016/j.ultrasmedbio.2022.08.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。