超声造影剂微泡稳定性用于膀胱压测量
本研究评估在典型膀胱测压(CMG)过程中,超声造影微泡(MB)使用次谐波辅助压力估计技术时的稳定性。通过粒径测量和超声造影成像实验,检验了静脉袋空气顶隙、MB稀释倍数和CMG灌注系统的影响。结果表明空气顶隙无显著影响,较高MB剂量更稳定,而CMG灌注系统对MB稳定性有不利影响。
要点
- CMG灌注系统会降低微泡稳定性,需关注灌注装置的影响。
- 较高微泡剂量可提高稳定性,但受FDA法规和声影伪影限制。
- 较低灌注速率可能有助于维持微泡稳定性。
局限性
- 研究为体外实验,未在临床患者中验证。
- 未报告样本量、具体设备参数和统计细节。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The goal of this study was to evaluate ultrasound contrast microbubbles (MB) stability during a typical cystometrogram (CMG) for bladder pressure measurement application using the subharmonic-aided pressure estimation technique. A detailed study of MB stability was required given two unique characteristics of this application: first, bulk infusion of MBs into the bladder through the CMG infusion system, and second, duration of a typical CMG which may last up to 30 min. To do so, a series of size measurement and contrast-enhanced ultrasound imaging studies under different conditions were performed and the effects of variables that we hypothesized have an effect on MB stability, namely, i) IV bag air headspace, ii) MB dilution factor, and iii) CMG infusion system were investigated. The results verified that air volume in intravenous (IV) bag headspace was not enough to have a significant effect on MB stability during a CMG. We also showed that higher MB dosage results in a more stable condition. Finally, the results indicated that the CMG infusion system adversely affects MB stability. In summary, to ensure MB stability during the entire duration of a CMG, lower filling rates (limited by estimated bladder capacity in clinical applications) and/or higher MB dosage (limited by FDA regulations and shadowing artifact) and/or the consideration of alternative catheter design may be needed.
引用
- Kalayeh, Fowlkes, Claflin, et al. Ultrasound Contrast Stability for Urinary Bladder Pressure Measurement. Ultrasound Med Biol. 2022. PMID: 36244919. doi:10.1016/j.ultrasmedbio.2022.08.008
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。