沸腾组织碎化术用于离体人前列腺组织机械消融的初步评估
本研究以快速尸检获取的24例新鲜离体人前列腺组织为对象,评估沸腾组织碎化术(BH)行机械消融的可行性,并以剪切波弹性成像确认组织机械属性与临床代表范围相符。BH在B模式下焦点呈高回声、能量多普勒可检出气泡,随治疗进展出现低回声区提示组织碎裂;1 ms脉冲较10 ms脉冲治疗速度快一倍,组织学显示病灶内细胞碎屑完全均质化。结论支持BH作为前列腺消融临床转化选项的进一步研究。
要点
- BH以毫秒级脉冲冲击波诱导机械消融,B模式下焦点高回声、多普勒可检出气泡,进展后出现低回声区提示组织碎裂。
- 离体前列腺组织剪切波弹性成像为37.9±22.2 kPa,处于临床典型范围内,支持该模型代表性。
- 1 ms脉冲较10 ms脉冲BH治疗速度快一倍,组织学证实病灶内细胞碎屑完全均质化。
关键影像表现/征象
- BH过程中B模式下焦点可见高回声区,能量多普勒可检出BH诱导的气泡。
- 随治疗进展,组织出现低回声区,提示组织碎裂成功。
- 组织学分析显示病灶内含完全均质化的细胞碎屑,符合组织碎化术诱导的机械消融。
诊断标准与关键数值
- 离体前列腺组织剪切波弹性成像为37.9±22.2 kPa,处于临床观察到的典型范围内。
- 1 ms脉冲较10 ms脉冲的BH治疗速度快一倍。
临床应用与阅片要点
- BH消融时可在B模式关注焦点高回声及后续低回声区,并可用能量多普勒检出气泡作为治疗过程参考。
- 脉冲宽度影响治疗速度,1 ms脉冲较10 ms脉冲快一倍,参数选择需权衡效率与组织碎裂效果。
局限性
- 研究为离体人前列腺组织可行性评估,非在体临床研究,尚不能直接外推至临床疗效与安全性。
- 样本经快速尸检获取,虽经剪切波弹性成像确认机械属性在临床典型范围内,仍与在体组织存在差异。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Boiling histotripsy (BH) is a focused ultrasound technology that uses millisecond-long pulses with shock fronts to induce mechanical tissue ablation. The pulsing scheme and mechanisms of BH differ from those of cavitation cloud histotripsy, which was previously developed for benign prostatic hyperplasia. The goal of the work described here was to evaluate the feasibility of using BH to ablate fresh ex vivo human prostate tissue as a proof of principle for developing BH for prostate applications. Fresh human prostate samples (N = 24) were obtained via rapid autopsy (<24 h after death, institutional review board exempt). Samples were analyzed using shear wave elastography to ensure that mechanical properties of autopsy tissue were clinically representative. Samples were exposed to BH using 10- or 1-ms pulses with 1% duty cycle under real-time B-mode and Doppler imaging. Volumetric lesions were created by sonicating 1-4 rectangular planes spaced 1 mm apart, containing a grid of foci spaced 1-2 mm apart. Tissue then was evaluated grossly and histologically, and the lesion content was analyzed using transmission electron microscopy and scanning electron microscopy. Observed shear wave elastography characterization of ex vivo prostate tissue (37.9 ± 22.2 kPa) was within the typical range observed clinically. During BH, hyperechoic regions were visualized at the focus on B-mode, and BH-induced bubbles were also detected using power Doppler. As treatment progressed, hypoechoic regions of tissue appeared, suggesting successful tissue fractionation. BH treatment was twofold faster using shorter pulses (1 ms vs. 10 ms). Histological analysis revealed lesions containing completely homogenized cell debris, consistent with histotripsy-induced mechanical ablation. It was therefore determined that BH is feasible in fresh ex vivo human prostate tissue producing desired mechanical ablation. The study supports further work aimed at translating BH technology as a clinical option for prostate ablation.
引用
- Khokhlova, Rosnitskiy, Tsysar, et al. Initial Assessment of Boiling Histotripsy for Mechanical Ablation of Ex Vivo Human Prostate Tissue. Ultrasound Med Biol. 2022. PMID: 36207225. doi:10.1016/j.ultrasmedbio.2022.07.014
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。