1.5-T与3.0-T MRI肝脏铁定量多中心可重复性研究
该多中心前瞻性横断面研究纳入207例已知或疑似肝铁过载参与者,在同一天采用1.5 T和2.89 T或3.0 T多回波梯度回波MRI进行R2* mapping,以标准化1.5 T自旋回波R2 MRI为参考。结果显示R2*与LIC呈线性关系,同场强内校准高度可重复(P > .43),R2*检测临床相关LIC阈值的AUC > 0.98。
要点
- R2* MRI可在临床相关LIC范围内准确、可重复地定量肝铁过载。
- 同场强内不同中心校准无显著差异(P > .43),支持跨厂商应用。
- R2*检测临床相关LIC阈值的AUC > 0.98,诊断性能高。
核心结论
- R2* MRI可在临床相关肝铁浓度范围内实现准确且可重复的肝铁过载定量。
- 研究提供的校准关系支持R2*法肝铁定量在临床广泛推广。
研究设计与样本
- 前瞻性横断面研究,纳入207例已知或疑似肝铁过载参与者(平均年龄38岁±20,男性117例),2015年3月至2019年9月。
- 四家学术医学中心参与,使用三个不同厂商的MRI扫描仪(三台1.5 T、一台2.89 T、两台3.0 T)。
- 同日行1.5 T和2.89 T或3.0 T多回波梯度回波MRI R2* mapping,图像在单中心重建分析。
- 参考LIC采用标准化1.5 T自旋回波商用R2 MRI方法测量。
诊断标准与关键数值
- 1.5 T校准公式:LIC(mg/g干重)= -0.16 + 2.603 × 10^-2 R2*(s^-1)。
- 2.89 T校准公式:LIC(mg/g)= -0.03 + 1.400 × 10^-2 R2*(s^-1)。
- 3.0 T校准公式:LIC(mg/g)= -0.03 + 1.349 × 10^-2 R2*(s^-1)。
- 同场强内所有校准高度可重复,无显著中心特异性差异(所有比较P > .43)。
- R2*检测临床相关LIC阈值的AUC > 0.98。
临床应用与阅片要点
- 阅片时可直接应用本研究的场强特异性校准公式将R2*值转换为LIC。
- 不同厂商同场强扫描仪间校准可重复,支持跨中心标准化报告。
局限性
- 研究为前瞻性横断面设计,参考标准为商用R2 MRI而非活检。
- 仅纳入四家中心、三个厂商设备,校准公式的广泛适用性需进一步验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background MRI is a standard of care tool to measure liver iron concentration (LIC). Compared with regulatory-approved R2 MRI, R2* MRI has superior speed and is available in most MRI scanners; however, the cross-vendor reproducibility of R2*-based LIC estimation remains unknown. Purpose To evaluate the reproducibility of LIC via single-breath-hold R2* MRI at both 1.5 T and 3.0 T with use of a multicenter, multivendor study. Materials and Methods Four academic medical centers using MRI scanners from three different vendors (three 1.5-T scanners, one 2.89-T scanner, and two 3.0-T scanners) participated in this prospective cross-sectional study. Participants with known or suspected liver iron overload were recruited to undergo multiecho gradient-echo MRI for R2* mapping at 1.5 T and 3.0 T (2.89 T or 3.0 T) on the same day. R2* maps were reconstructed from the multiecho images and analyzed at a single center. Reference LIC measurements were obtained with a commercial R2 MRI method performed using standardized 1.5-T spin-echo imaging. R2*-versus-LIC calibrations were generated across centers and field strengths using linear regression and compared using F tests. Receiver operating characteristic (ROC) curve analysis was used to determine the diagnostic performance of R2* MRI in the detection of clinically relevant LIC thresholds. Results A total of 207 participants (mean age, 38 years ± 20 [SD]; 117 male participants) were evaluated between March 2015 and September 2019. A linear relationship was confirmed between R2* and LIC. All calibrations within the same field strength were highly reproducible, showing no evidence of statistically significant center-specific differences (P > .43 across all comparisons). Calibrations for 1.5 T and 3.0 T were generated, as follows: for 1.5 T, LIC (in milligrams per gram [dry weight]) = -0.16 + 2.603 × 10-2 R2* (in seconds-1); for 2.89 T, LIC (in milligrams per gram) = -0.03 + 1.400 × 10-2 R2* (in seconds-1); for 3.0 T, LIC (in milligrams per gram) = -0.03 + 1.349 × 10-2 R2* (in seconds-1). Liver R2* had high diagnostic performance in the detection of clinically relevant LIC thresholds (area under the ROC curve, >0.98). Conclusion R2* MRI enabled accurate and reproducible quantification of liver iron overload over clinically relevant ranges of liver iron concentration (LIC). The data generated in this study provide the necessary calibrations for broad clinical dissemination of R2*-based LIC quantification. ClinicalTrials.gov registration no.: NCT02025543 © RSNA, 2022 Online supplemental material is available for this article.