现代低场MRI在肌肉骨骼系统的实践考量、机遇与挑战
本文综述了现代低场(0.55T)全身MRI系统在肌肉骨骼影像中的应用。该系统具有较低的总拥有成本,并通过现代信号链、先进射频线圈和深度学习重建改善图像质量。文章讨论了0.55T下肌肉骨骼组织对比度的变化、脂肪抑制和金属伪影减少等技术的可行性,以及宽体架设计为介入MRI带来的机遇。
要点
- 0.55T低场MRI系统成本更低,适合推广普及。
- 现代技术可改善低场信噪比和图像质量。
- 低场下金属伪影减少效果显著,利于植入物评估。
关键影像表现/征象
- 肌肉骨骼组织对比度在0.55T下表现不同,需仔细设计脉冲序列。
- 化学位移选择脂肪抑制在现代低场MRI中可靠且均匀。
- Dixon梯度回波序列可提供多对比度选项,包括增强后成像。
- 低场强有利于金属伪影减少,包括切片编码金属伪影校正技术。
临床应用与阅片要点
- 可利用宽体架设计开展介入MRI操作。
- 优先选用Dixon梯度回波序列实现多对比度成像。
- 使用化学位移选择法进行脂肪抑制以保证均匀性。
- 对高敏感性金属植入物可采用切片编码金属伪影校正。
局限性
- 信号噪声比随场强降低,需依赖先进重建算法补偿。
- 低场下脂肪抑制本身更具挑战性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Magnetic resonance imaging (MRI) provides essential information for diagnosing and treating musculoskeletal disorders. Although most musculoskeletal MRI examinations are performed at 1.5 and 3.0 T, modern low-field MRI systems offer new opportunities for affordable MRI worldwide. In 2021, a 0.55 T modern low-field, whole-body MRI system with an 80-cm-wide bore was introduced for clinical use in the United States and Europe. Compared with current higher-field-strength MRI systems, the 0.55 T MRI system has a lower total ownership cost, including purchase price, installation, and maintenance. Although signal-to-noise ratios scale with field strength, modern signal transmission and receiver chains improve signal yield compared with older low-field magnetic resonance scanner generations. Advanced radiofrequency coils permit short echo spacing and overall compacter echo trains than previously possible. Deep learning-based advanced image reconstruction algorithms provide substantial improvements in perceived signal-to-noise ratios, contrast, and spatial resolution. Musculoskeletal tissue contrast evolutions behave differently at 0.55 T, which requires careful consideration when designing pulse sequences. Similar to other field strengths, parallel imaging and simultaneous multislice acquisition techniques are vital for efficient musculoskeletal MRI acquisitions. Pliable receiver coils with a more cost-effective design offer a path to more affordable surface coils and improve image quality. Whereas fat suppression is inherently more challenging at lower field strengths, chemical shift selective fat suppression is reliable and homogeneous with modern low-field MRI technology. Dixon-based gradient echo pulse sequences provide efficient and reliable multicontrast options, including postcontrast MRI. Metal artifact reduction MRI benefits substantially from the lower field strength, including slice encoding for metal artifact correction for effective metal artifact reduction of high-susceptibility metallic implants. Wide-bore scanner designs offer exciting opportunities for interventional MRI. This review provides an overview of the economical aspects, signal and image quality considerations, technological components and coils, musculoskeletal tissue relaxation times, and image contrast of modern low-field MRI and discusses the mainstream and new applications, challenges, and opportunities of musculoskeletal MRI.
引用
- Khodarahmi, Keerthivasan, Brinkmann, et al. Modern Low-Field MRI of the Musculoskeletal System: Practice Considerations, Opportunities, and Challenges. Invest Radiol. 2022. PMID: 36165841. doi:10.1097/RLI.0000000000000912
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。