相敏乳腺断层摄影性能的初步研究

A Pilot Study to Assess the Performance of Phase-Sensitive Breast Tomosynthesis.
PubMed 文献 · Radiology · 2022-09-27

一项前瞻性初步研究纳入50例患者52个病灶,由9名放射科医生独立评估相敏乳腺断层摄影(PBT)与常规数字乳腺断层摄影(DBT)的图像质量和病灶良恶性分类性能。结果显示DBT图像质量评分更高(3.8 vs 2.9),但两者诊断良恶性的AUC均为0.74;PBT的平均辐射剂量较DBT降低24%(1.78 mGy vs 2.34 mGy)。

要点

  • PBT与DBT对乳腺病灶良恶性分类的AUC均为0.74,诊断性能相当。
  • PBT平均辐射剂量较DBT降低24%,但图像质量主观评分较低。
  • PBT的诊断性能尚需更大规模研究确定。

关键影像表现/征象

  • DBT图像外观和一般特征可见性的总平均图像质量评分(3.8)高于PBT(2.9),每项比较P < .002。
  • PBT图像采集的平均辐射剂量为1.78 mGy,DBT为2.34 mGy,平均降低24%(P < .001)。

诊断标准与关键数值

  • 图像质量采用1至5分Likert量表主观评分。
  • 病灶良恶性分类性能用受试者工作特征曲线下面积(AUC)比较,PBT和DBT的AUC均为0.74。

临床应用与阅片要点

  • PBT可在降低辐射剂量的同时保持与DBT相当的病灶良恶性分类性能,但图像质量主观评分较低,阅片时需注意图像外观和特征可见性的差异。

局限性

  • 初步研究样本量小(50例患者52个病灶),诊断性能需更大规模研究确定。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background A new modality, phase-sensitive breast tomosynthesis (PBT), may have similar diagnostic performance to conventional breast tomosynthesis but with a reduced radiation dose. Purpose To perform a pilot study of the performance of a novel PBT system compared with conventional digital breast tomosynthesis (DBT) in patients undergoing additional diagnostic imaging workup for breast lesions. Materials and Methods In a prospective study from June 2020 to March 2021, participants with suspicious breast lesions detected at screening DBT or MRI were recruited for additional PBT imaging before additional diagnostic workup or biopsy. In this pilot study, nine radiologists independently evaluated image quality and assessed the likelihood of lesion malignancy by retrospectively evaluating DBT and PBT images in two separate reading sessions. Image quality was rated subjectively using a Likert scale from 1 to 5. Areas under the receiver operating characteristic curve (AUCs) were used to compare the lesion classification (malignant vs benign) performance of the radiologists. Results Images in 50 patients (mean age, 56 years ± 12 [SD]; 49 women) with 52 evaluable lesions (28 malignant) were assessed. For image appearance and general feature visibility, DBT images had a higher total mean image quality score (3.8) than PBT images (2.9), with P < .002 for each comparison. For classification of lesions as benign or malignant, the AUCs were 0.74 for both PBT and DBT. PBT images were acquired at a 24% mean radiation dose reduction (mean, 1.78 mGy vs 2.34 mGy for DBT; P < .001). Conclusion The phase-sensitive breast tomosynthesis system had a 24% lower mean radiation dose compared with digital breast tomosynthesis, although with lower image quality. Diagnostic performance of the system remains to be determined in larger studies. © RSNA, 2022 Online supplemental material is available for this article. See also the editorial by Gao and Moy in this issue.

来源:Radiology,查看原文。版权归原作者所有。

引用

  1. Fajardo, Hillis, Zheng, et al. A Pilot Study to Assess the Performance of Phase-Sensitive Breast Tomosynthesis. Radiology. 2022. PMID: 36165790. doi:10.1148/radiol.213198

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。