深度学习CT重建对小低衰减病灶可检测性的多读者体模研究
该研究采用双体模结构和24名不同经验读者,以两选一强制选择法评估深度学习图像重建(DLIR)与滤波反投影(FBP)及迭代重建(IR)在低对比可检测性和辐射节省潜力方面的表现。结果显示,在常规辐射水平下,DLIR中、高设置相较于FBP和IR在辐射降低达90%时仍保持非劣效,而IR仅在辐射降低30%和50%时非劣效于常规辐射FBP。读者经验与诊断准确性无相关性。
要点
- DLIR高设置可在辐射降低70%时保持与常规剂量FBP相当的低对比可检测性。
- DLIR中、高设置在辐射降低达90%时仍非劣效于常规剂量FBP和IR。
- 读者经验水平不影响低对比病灶检测准确性。
局限性
- 体模研究,结果可能不能直接推广到临床患者。
- 仅评估了低对比低衰减病灶,未涵盖其他类型病灶。
- 读者经验与诊断准确性无相关性(R2=0.005),但样本量有限。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND. Iterative reconstruction (IR) techniques are susceptible to contrast-dependent spatial resolution, limiting overall radiation dose reduction potential. Deep learning image reconstruction (DLIR) may mitigate this limitation. OBJECTIVE. The purpose of our study was to evaluate low-contrast detectability performance and radiation-saving potential of a DLIR algorithm in comparison with filtered back projection (FBP) and IR using a human multireader noninferiority study design and task-based observer modeling. METHODS. A dual-phantom construct, consisting of a low-contrast detectability module (21 low-contrast hypoattenuating objects in seven sizes [2.4-10.0 mm] and three contrast levels [-15, -10, -5 HU] embedded within liver-equivalent background) and a phantom, was imaged at five radiation exposures (CTDIvol range, 1.4-14.0 mGy; size-specific dose estimate, 2.5-25.0 mGy; 90%-, 70%-, 50%-, and 30%-reduced radiation levels and full radiation level) using an MDCT scanner. Images were reconstructed using FBP, hybrid IR (ASiR-V), and DLIR (TrueFidelity). Twenty-four readers of varying experience levels evaluated images using a two-alternative forced choice. A task-based observer model (detectability index [d']) was calculated. Reader performance was estimated by calculating the AUC using a noninferiority method. RESULTS. Compared with FBP and IR methods at routine radiation levels, DLIR medium and DLIR high settings showed noninferior performance through a 90% radiation reduction (except DLIR medium setting at 70% reduced level). The IR method was non-inferior to routine radiation FBP only for 30% and 50% radiation reductions. No significant difference in d' was observed between routine radiation FBP and DLIR high setting through a 70% radiation reduction. Reader experience was not correlated with diagnostic accuracy (R2 = 0.005). CONCLUSION. Compared with FBP or IR methods at routine radiation levels, certain DLIR algorithm weightings yielded noninferior low-contrast detectability with radiation reductions of up to 90% as measured by 24 human readers and up to 70% as assessed by a task-based observer model. CLINICAL IMPACT. DLIR has substantial potential to preserve contrast-dependent spatial resolution for the detection of hypoattenuating lesions at decreased radiation levels in a phantom model, addressing a major shortcoming of current IR techniques.
引用
- Toia, Zamora, Singleton, et al. Detectability of Small Low-Attenuation Lesions With Deep Learning CT Image Reconstruction: A 24-Reader Phantom Study. AJR Am J Roentgenol. 2022. PMID: 36129222. doi:10.2214/AJR.22.28407
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。