深度学习图像重建提升腹部双能CT碘定量准确性与图像质量:体模与临床研究
本研究在体模和腹部临床患者中比较深度学习图像重建(DLIR)与其他重建算法对双能CT碘定量准确性和图像质量的影响。结论提示DLIR可降低图像噪声和碘浓度变异性,而不影响碘定量准确性。
要点
- DLIR不改变双能CT碘定量准确性,各算法测得碘浓度一致。
- DLIR较其他重建算法显著降低图像噪声和碘浓度变异系数。
- 在腹部动脉期和门静脉期图像中,DLIR均能改善图像质量。
关键影像表现/征象
- 体模实验中各重建算法测得的碘浓度均在标称值±8%以内,均方根偏差为0.08-0.36 mgI/mL,不受辐射剂量影响。
诊断标准与关键数值
- 体模碘浓度准确性:各算法测得值在标称值±8%以内,均方根偏差0.08-0.36 mgI/mL。
临床应用与阅片要点
- DLIR可用于腹部双能CT,在保持碘定量准确性的同时降低图像噪声,有助于提高图像质量和碘浓度测量的可重复性。
- 阅片时应注意DLIR可能改变图像纹理,但碘浓度值不受影响,可放心用于定量分析。
局限性
- 临床研究为回顾性设计,样本量有限(50例),且仅使用单一厂商的快速千伏切换双能CT设备。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the effect of deep learning image reconstruction (DLIR) on the accuracy of iodine quantification and image quality of dual-energy CT (DECT) compared to that of other reconstruction algorithms in a phantom experiment and an abdominal clinical study.
METHODS: An elliptical phantom with five different iodine concentrations (1-12 mgI/mL) was imaged five times with fast-kilovoltage-switching DECT for three target volume CT dose indexes. All images were reconstructed using filtered back-projection, iterative reconstruction (two levels), and DLIR algorithms. Measured and nominal iodine concentrations were compared among the algorithms. Contrast-enhanced CT of the abdomen with the same scanner was acquired in clinical patients. In arterial and portal venous phase images, iodine concentration, image noise, and coefficients of variation for four locations were retrospectively compared among the algorithms. One-way repeated-measures analyses of variance were used to evaluate differences in the iodine concentrations, standard deviations, coefficients of variation, and percentages of error among the algorithms.
RESULTS: In the phantom study, the measured iodine concentrations were equivalent among the algorithms: within ± 8% of the nominal values, with root-mean-square deviations of 0.08-0.36 mgI/mL, regardless of radiation dose. In the clinical study (50 patients; 35 men; mean age, 68 ± 11 years), iodine concentrations were equivalent among the algorithms for each location (all p > .99). Image noise and coefficients of variation were lower with DLIR than with the other algorithms (all p < .01).
CONCLUSIONS: The DLIR algorithm reduced image noise and variability of iodine concentration values compared with other reconstruction algorithms in the fast-kilovoltage-switching dual-energy CT.
KEY POINTS: • In the phantom study, standard deviations and coefficients of variation in iodine quantification were lower on images with the deep learning image reconstruction algorithm than on those with other algorithms. • In the clinical study, iodine concentrations of measurement location in the upper abdomen were consistent across four reconstruction algorithms, while image noise and variability of iodine concentrations were lower on images with the deep learning image reconstruction algorithm.
引用
- Fukutomi, Sofue, Ueshima, et al. Deep learning image reconstruction to improve accuracy of iodine quantification and image quality in dual-energy CT of the abdomen: a phantom and clinical study. Eur Radiol. 2022. PMID: 36114848. doi:10.1007/s00330-022-09127-1