膀胱体模中对比增强超声测压:通向无导管排尿性膀胱测压图

Pressure Measurement in a Bladder Phantom Using Contrast-Enhanced Ultrasonography-A Path to a Catheter-Free Voiding Cystometrogram.
PubMed 文献 · Invest Radiol · 2022-08-28

本研究在膀胱体模中验证次谐波辅助压力估计(SHAPE)技术测量膀胱压力的可行性。采用脂质外壳微泡灌注体模,模拟膀胱测压图充盈及充盈后事件并加压,同步采集次谐波信号。结果显示SHAPE信号与体模压力呈强负线性相关(r²>0.9,P<0.001),但转换因子随事件和时间变化,提示需按不同时间点调整转换因子。

要点

  • SHAPE信号与膀胱体模压力呈强负线性相关,r²>0.9,P<0.001。
  • SHAPE转换因子在充盈与充盈后事件间及不同时间点存在变化。
  • 微泡次谐波发射可作为膀胱压力变化的良好指标,支持无导管测压方向。

核心结论

  • 微泡次谐波发射是膀胱体模压力变化的优秀指标,SHAPE信号与体模压力强相关,表明该方法可用于膀胱测压图中测量膀胱压力。
  • 不同测压图事件(不同时间点)可能需要不同的SHAPE转换因子,需进一步优化协议以引入人体研究。

研究设计与样本

  • 研究类型为体外膀胱体模实验,使用实验室自制脂质外壳微泡(成分类似DEFINITY)稀释于0.9%生理盐水,经测压图输注系统注入体模。
  • 模拟测压图包括1个充盈事件和4个充盈后事件,在每个事件中按小、中、大临床相关水平多次加压,同步采集微泡次谐波信号。
  • 采用基于小时间窗的信号处理技术校正模拟测压图过程中微泡次谐波信号的时间衰减,线性拟合确定SHAPE转换因子(斜率)。

关键影像表现/征象

  • 每个充盈及充盈后事件中,SHAPE与膀胱体模压力均呈强负线性关系(r²>0.9,均方根误差<0.3 dB,标准误<0.01 dB,P<0.001)。
  • SHAPE在测量体模压力时表现出瞬态行为,转换因子(dB/cm H₂O)随充盈与充盈后事件及充盈后时间变化。
  • 转换因子幅度在充盈后立即趋于增加,随后随时间下降。

诊断标准与关键数值

  • SHAPE与膀胱体模压力的线性拟合参数:r²>0.9,均方根误差<0.3 dB,标准误<0.01 dB,P<0.001。
  • SHAPE转换因子单位为dB/cm H₂O,其值在充盈与充盈后事件间及不同充盈后时间点存在变化。

临床应用与阅片要点

  • 阅片或分析时需注意SHAPE信号与压力呈负线性关系,且转换因子随测压图事件和时间变化,不宜固定使用单一转换因子。
  • 信号处理需采用小时间窗校正微泡次谐波信号的时间衰减,以避免压力估计偏差。

局限性

  • 研究在体外膀胱体模中进行,非人体在体研究,结果外推至临床需进一步验证。
  • SHAPE转换因子随事件和时间变化,需优化协议后才能引入人体受试者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The long-term goal of this study is to investigate the efficacy of a novel, ultrasound-based technique called subharmonic-aided pressure estimation (SHAPE) to measure bladder pressure as a part of a cystometrogram (CMG) in a urodynamic test (ie, pressure-flow study). SHAPE is based on the principle that subharmonic emissions from ultrasound contrast microbubbles (MBs) decrease linearly with an increase in ambient pressure. We hypothesize that, using the SHAPE technique, we can measure voiding bladder pressure catheter-free. This is of importance because the CMG catheter, due to its space-occupying property and non-physiological effects, can undermine the reliability of the test during voiding and cause misdiagnosis. In this study, we tested this hypothesis and optimized the protocol in a controlled benchtop environment.

MATERIALS AND METHODS: A bladder phantom was designed and built, capable of simulating clinically relevant bladder pressures. Laboratory-made lipid-shelled MBs (similar in composition to the commercial agent, DEFINITY) was diluted in 0.9% normal saline and infused into the bladder phantom using the CMG infusion system. A typical simulated CMG consists of 1 filling and 4 post-filling events. During CMG events, the bladder phantom is pressurized multiple times at different clinically relevant levels (small, medium, and large) to simulate bladder pressures. Simultaneous with pressurization, MB subharmonic signal was acquired. For each event, the change in MB subharmonic amplitude was correlated linearly with the change in bladder phantom pressure, and the SHAPE conversion factor (slope of the linear fit) was determined. In doing so, a specific signal processing technique (based on a small temporal window) was used to account for time-decay of MB subharmonic signal during a simulated CMG.

RESULTS: A strong inverse linear relationship was found to exist between SHAPE and bladder phantom pressures for each of the CMG filling and post-filling events ( r2> 0.9, root mean square error < 0.3 dB, standard error <0.01 dB, and P < 0.001). SHAPE showed a transient behavior in measuring bladder phantom pressure. The SHAPE conversion factor (in dB/cm H 2 O) varied between filling and post-filling events, as well as by post-filling time. The magnitude of the SHAPE conversion factor tended to increase immediately after filling and then decreases with time.

CONCLUSIONS: Microbubble subharmonic emission is an excellent indicator of bladder phantom pressure variation. The strong correlation between SHAPE signal and bladder phantom pressure is indicative of the applicability of this method in measuring bladder pressure during a CMG. Our results suggest that different SHAPE conversion factors may be needed for different events during a CMG (ie, at different time points of a CMG). These findings will help us better protocolize this method for introduction into human subjects and allow us to take the next step toward developing a catheter-free voiding CMG using SHAPE.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Kalayeh, Fowlkes, Chen, et al. Pressure Measurement in a Bladder Phantom Using Contrast-Enhanced Ultrasonography-A Path to a Catheter-Free Voiding Cystometrogram. Invest Radiol. 2022. PMID: 36070543. doi:10.1097/RLI.0000000000000919

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。