本研究开发并验证了一种基于深度学习的算法,用于自动检测和分级钆塞酸增强肝脏MRI动脉期运动相关伪影。该算法在内部和外部验证集中均表现出较高的敏感度和特异度,并能快速完成自动分级。

要点

  • 深度学习算法可自动检测动脉期运动伪影,内部验证敏感度91.5%、特异度96.8%。
  • 外部验证显示算法泛化能力良好,敏感度93.3%、特异度91.6%。
  • 每例检查自动分级耗时约4.11-4.22秒,有望辅助临床工作流程。

局限性

  • 研究为回顾性设计,训练和内部验证数据来自单一机构,外部验证虽涉及多机构但样本量有限。
  • 仅针对钆塞酸增强肝脏MRI动脉期,未评估其他序列或对比剂。
  • 伪影分级基于人工评估的3级系统,可能存在主观性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to develop and validate a deep learning-based algorithm (DLA) for automatic detection and grading of motion-related artifacts on arterial phase liver magnetic resonance imaging (MRI).

MATERIALS AND METHODS: Multistep DLA for detection and grading of motion-related artifacts, based on the modified ResNet-101 and U-net, were trained using 336 arterial phase images of gadoxetic acid-enhanced liver MRI examinations obtained in 2017 (training dataset; mean age, 68.6 years [range, 18-95]; 254 men). Motion-related artifacts were evaluated in 4 different MRI slices using a 3-tier grading system. In the validation dataset, 313 images from the same institution obtained in 2018 (internal validation dataset; mean age, 67.2 years [range, 21-87]; 228 men) and 329 from 3 different institutions (external validation dataset; mean age, 64.0 years [range, 23-90]; 214 men) were included, and the per-slice and per-examination performances for the detection of motion-related artifacts were evaluated.

RESULTS: The per-slice sensitivity and specificity of the DLA for detecting grade 3 motion-related artifacts were 91.5% (97/106) and 96.8% (1134/1172) in the internal validation dataset and 93.3% (265/284) and 91.6% (948/1035) in the external validation dataset. The per-examination sensitivity and specificity were 92.0% (23/25) and 99.7% (287/288) in the internal validation dataset and 90.0% (72/80) and 96.0% (239/249) in the external validation dataset, respectively. The processing time of the DLA for automatic grading of motion-related artifacts was from 4.11 to 4.22 seconds per MRI examination.

CONCLUSIONS: The DLA enabled automatic and instant detection and grading of motion-related artifacts on arterial phase gadoxetic acid-enhanced liver MRI.