0.55 T MRI 对颅内动脉瘤的诊断效能

Diagnostic Performance of 0.55 T MRI for Intracranial Aneurysm Detection.
PubMed 文献 · Invest Radiol · 2022-09-02

这项前瞻性研究纳入17例疑似颅内动脉瘤患者,在DSA前1天行0.55 T TOF MRA,并与DSA对照。16例患者共19个动脉瘤被两种方法一致识别;0.55 T与DSA在动脉瘤整体大小、宽度和瘤颈测量上无显著差异,高度测量略大;盲法阅片者用0.55 T图像正确检出全部动脉瘤。

要点

  • 0.55 T TOF MRA可可靠检出颅内动脉瘤,盲法阅片全部正确。
  • 0.55 T与DSA测量动脉瘤整体大小、宽度、瘤颈无显著差异,高度略偏大。
  • 0.55 T与1.5/3 T TOF MRA及阅片者间一致性均极佳(ICC>0.94)。

核心结论

  • 现代0.55 T MRI采集的TOF-MRA是检出和初步评估颅内动脉瘤的可靠工具。

研究设计与样本

  • 前瞻性研究,17例疑似单发或多发颅内动脉瘤患者,在DSA前1天行0.55 T TOF MRA。
  • 两名神经放射科医生独立在0.55 T、1.5 T和3 T 3D-TOF MRA源图像及2D/3D旋转血管造影上测量动脉瘤瘤颈、宽度和高度。
  • 主要分析用Bland-Altman图、Wilcoxon检验和组内相关系数(ICC)评估0.55 T TOF MRA与DSA的跨模态一致性。
  • 第三名对病史不知情的神经放射科医生盲法筛查17例患者及15例额外健康者的0.55 T TOF MRA数据集,判断动脉瘤有无及位置。

诊断标准与关键数值

  • 0.55 T TOF MRA与DSA总体动脉瘤大小(高度/宽度/瘤颈均值)无显著差异(P=0.178)。
  • 0.55 T TOF MRA与DSA平均宽度(P=0.778)和瘤颈值(P=0.190)无显著差异。
  • 0.55 T TOF MRA平均高度显著大于DSA(P=0.020)。
  • 0.55 T与1.5/3 T TOF MRA跨模态一致性及阅片者间一致性均极佳(ICC>0.94)。
  • 盲法阅片者使用0.55 T图像在32例患者数据集中正确检出全部动脉瘤。

临床应用与阅片要点

  • 0.55 T TOF MRA可用于颅内动脉瘤的检出和初步评估,测量整体大小、宽度和瘤颈时与DSA一致性较好。
  • 测量动脉瘤高度时需注意0.55 T TOF MRA可能较DSA偏大,报告时应结合其他序列或DSA确认。

局限性

  • 样本量较小(17例疑似患者加15例健康者,共19个动脉瘤),结果需更大样本验证。
  • 为单中心前瞻性研究,仅评估0.55 T TOF MRA,未与其他低场强设备比较。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Intracranial aneurysm (IA) is the main cause of subarachnoid hemorrhages. Time-of-flight (TOF) magnetic resonance angiography (MRA) at 1.5 T or 3 T magnetic resonance imaging (MRI) is a well-established method for the diagnosis of IA. The aim of this prospective study was to evaluate the performance of a modern 0.55 T MRI in the diagnosis of IAs in comparison to digital subtraction angiography (DSA) as a standard of reference.

MATERIALS AND METHODS: Seventeen patients with suspicion of single or multiple IAs underwent TOF MRA at 0.55 T MRI 1 day before DSA. Two neuroradiologists independently measured the aneurysm neck, width, and height on 0.55 T, 1.5 T, and 3 T 3D-TOF MRA source images and 2D/3D rotational angiography. The main analysis assessed the intermodality agreement between 0.55 T TOF MRA and DSA using Bland-Altman plots, a Wilcoxon test, and the intraclass correlation coefficient (ICC). In a secondary analysis, aneurysm dimensions were compared between 0.55 T TOF MRA and 1.5/3 T TOF MRA. Interreader agreement was evaluated by ICC. A third neuroradiologist blinded to patient history screened 0.55 T TOF MRA data sets of the aforementioned 17 patients and 15 additional healthy patients for the presence and location of aneurysms.

RESULTS: A total of 19 aneurysms in 16 patients were identified in both 0.55 T MRA and DSA. Measurements of the 2 nonblinded readers showed no significant differences between 0.55 T TOF MRA and DSA in the overall aneurysm size (calculated as the mean from height/width/neck) ( P = 0.178), as well as in the mean width ( P = 0.778) and neck values ( P = 0.190). The mean height was significantly larger in 0.55 T TOF MRA in comparison to DSA ( P = 0.020). Intermodality (1.5/3 T TOF MRA) and interrater agreement were excellent (ICC > 0.94). Of the 32 data sets of patients with and without IA, the blinded reader detected all aneurysms correctly by using 0.55 T images.

CONCLUSIONS: TOF-MRA acquired with a modern 0.55 T MRI is a reliable tool for the detection and initial assessment of IAs.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Osmanodja, Rösch, Knott, et al. Diagnostic Performance of 0.55 T MRI for Intracranial Aneurysm Detection. Invest Radiol. 2022. PMID: 36070538. doi:10.1097/RLI.0000000000000918

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。