肝脏MR指纹成像用于肝组织表征:组织病理学对照研究
23例接受肝活检的患者中,MRF衍生指标用于鉴别中重度与轻度/无组织学改变,其中脂肪变和炎症的诊断效能较高,纤维化和铁沉积相对较低。MRF可在单次屏气下同时定量T1、T2、T2*和质子密度脂肪分数(PDFF),与参考定量图高度一致且重复性良好。
要点
- MRF可在单次屏气下同时获得T1、T2、T2*和PDFF图,与参考定量MRI一致性高。
- MRF鉴别中重度脂肪变和炎症的AUC较高,纤维化和铁沉积的鉴别效能相对有限。
- MRF重复性良好,PDFF和T1的变异系数最低,适合纵向随访定量评估。
诊断标准与关键数值
- MRF与参考定量MRI的线性回归R²值:T1为0.86、T2为0.88、T2*为0.54、PDFF为0.99。
- 肝脏重复性总体变异系数:T1为3.2%、T2为5.5%、T2*为7.1%、PDFF为4.6%。
- MRF鉴别中重度与轻度/无改变AUC:纤维化0.62(95%CI:0.52,0.62)、炎症0.92(95%CI:0.88,0.92)、脂肪变0.97(95%CI:0.96,0.97)、铁沉积0.74(95%CI:0.57,0.74)。
临床应用与阅片要点
- 可利用单次屏气MRF同时获得T1、T2、T2*和PDFF定量图,减少多次扫描和屏气负担。
- PDFF与参考方法一致性极高(R²=0.99),适合作为脂肪定量的可靠指标。
- T2*与参考方法一致性相对较低(R²=0.54),解读铁沉积相关T2*值时需谨慎。
- MRF对炎症和脂肪变的鉴别效能较高,对纤维化和铁沉积的鉴别效能有限,报告时应结合临床和其他序列。
局限性
- 研究为单中心前瞻性设计,样本量有限,活检亚组仅23例。
- 采用二维单层MRF采集,未评估三维多层面或全肝覆盖的效能。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Background Liver MR fingerprinting (MRF) enables simultaneous quantification of T1, T2, T2*, and proton density fat fraction (PDFF) maps in single breath-hold acquisitions. Histopathologic correlation studies are desired for its clinical use. Purpose To compare liver MRF-derived metrics with separate reference quantitative MRI in participants with diffuse liver disease, evaluate scan-rescan repeatability of liver MRF, and validate MRF-derived measurements for histologic grading of liver biopsies. Materials and Methods This prospective study included participants with diffuse liver disease undergoing MRI from July 2021 to January 2022. Participants underwent two-dimensional single-section liver MRF and separate reference quantitative MRI. Linear regression, Bland-Altman plots, and coefficients of variation were used to assess the bias and repeatability of liver MRF measurements. For participants undergoing liver biopsy, the association between mapping and histologic grading was evaluated by using the Spearman correlation coefficient. Results Fifty-six participants (mean age, 59 years ± 15 [SD]; 32 women) were included to compare mapping techniques and 23 participants were evaluated with liver biopsy (mean age, 52.7 years ± 12.7; 14 women). The linearity of MRF with reference measurements in participants with diffuse liver disease (R2 value) for T1, T2, T2*, and PDFF maps was 0.86, 0.88, 0.54, and 0.99, respectively. The overall coefficients of variation for repeatability in the liver were 3.2%, 5.5%, 7.1%, and 4.6% for T1, T2, T2*, and PDFF maps, respectively. MRF-derived metrics showed high diagnostic performance in differentiating moderate or severe changes from mild or no changes (area under the receiver operating characteristic curve for fibrosis, inflammation, steatosis, and siderosis: 0.62 [95% CI: 0.52, 0.62], 0.92 [95% CI: 0.88, 0.92], 0.97 [95% CI: 0.96, 0.97], and 0.74 [95% CI: 0.57, 0.74], respectively). Conclusion Liver MR fingerprinting provided repeatable T1, T2, T2*, and proton density fat fraction maps in high agreement with reference quantitative mapping and may correlate with pathologic grades in participants with diffuse liver disease. © RSNA, 2022 Online supplemental material is available for this article.
引用
- Fujita, Sano, Cruz, et al. MR Fingerprinting for Liver Tissue Characterization: A Histopathologic Correlation Study. Radiology. 2022. PMID: 36040337. doi:10.1148/radiol.220736
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。