该回顾性研究纳入32例头颈部MRI扩散加权成像患者,比较常规SENSE(加速因子2)、快速SENSE(加速因子4)与EPICS(加速因子4)三种方法。结果显示EPICS较SENSE加速因子4提高对比噪声比,较常规扫描降低前后方向几何变形和变异系数,ADC值无显著差异。结论认为压缩SENSE结合高加速因子可在保持ADC值和扫描时间的同时改善头颈部DWI的图像质量、均匀性和变形。

要点

  • EPICS在头颈部DWI中较SENSE加速因子4提高各部位对比噪声比。
  • EPICS较常规SENSE扫描显著降低前后方向几何变形(p=0.02)。
  • EPICS较常规扫描显著降低变异系数,ADC值无显著差异。

关键影像表现/征象

  • EPICS在所有部位的对比噪声比均高于SENSE加速因子4。
  • EPICS在前后方向的几何变形显著低于常规扫描(p=0.02)。
  • EPICS的变异系数显著低于常规扫描[DWI:0.22(IQR:0.15-0.30)vs 0.32(IQR:0.24-0.40),p=0.02]。

诊断标准与关键数值

  • EPICS与常规扫描的ADC值比较无显著差异。

临床应用与阅片要点

  • 在头颈部DWI中可考虑使用EPICS以高加速因子扫描,在保持ADC值的同时改善图像均匀性和减少前后方向变形。

局限性

  • 回顾性研究,样本量较小(32例)。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: To evaluate diffusion-weighted imaging (DWI) using echo planar imaging (EPI) with compressed SENSE (EPICS) of the head and neck magnetic resonance imaging (MRI).

METHOD: We retrospectively observed 32 patients who underwent head and neck DWI according to either the conventional method (SENSE, reduction factor = 2), fast scanning method (SENSE, reduction factor = 4), or fast scanning method with EPICS (EPICS, reduction factor = 4). For quantitative analysis, contrast-to-noise-ratio (CNR), apparent diffusion coefficient (ADC) values, geometric distortion, and coefficient of variations (CV) were measured and compared. For qualitative analysis, all images were independently and blindly evaluated by two board-certified radiologists.

RESULTS: EPICS revealed the higher CNR between all location compared to those of SENSE with reduction factor = 4. Distortion in the anterior-posterior direction was significantly lower on EPICS than on the conventional scan (p = 0.02). A comparison between the ADC values of the EPICS and conventional scan revealed no significant differences. The CV was significantly lower for EPICS than the conventional scan [DWI: 0.22 (IQR: 0.15-0.30) vs 0.32 (IQR: 0.24-0.40), p = 0.02].

CONCLUSIONS: Compressed SENSE combined with the high acceleration factor can improve image quality, homogeneity, and distortion in the head and neck DWI maintaining ADC values and the scan time duration.