经颅多普勒超声质量保证用血流体模的研制

Development of a Flow Phantom for Transcranial Doppler Ultrasound Quality Assurance.
PubMed 文献 · Ultrasound Med Biol · 2022-08-26

本研究开发了一种大脑中动脉血流体模,用于在受控条件下比较不同经颅多普勒设备的流速估测。采用两种TCD设备与Zonare双功超声及定时集液金标准对比,Bland-Altman分析显示DWL系统测值在110 cm/s和210 cm/s时分别高于Spencer系统4.1 cm/s和12 cm/s。体模能系统量化设备间流速差异,对镰状细胞病卒中预防的流速阈值判读有临床意义。

要点

  • TCD设备间流速估测存在制造商相关的系统偏倚,高流速时差异更明显。
  • DWL Dopplerbox测值在210 cm/s时较Spencer ST3高约12 cm/s,接近卒中干预诊断阈值。
  • 使用TCD监测镰状细胞病时,需注意设备间偏倚对治疗决策的影响。

局限性

  • 仅比较了两种TCD设备,结论外推至其他品牌需谨慎。
  • 体模为自研非商用,未与临床在体测量直接对照验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Anecdotal evidence was recently brought to our attention suggesting a potential difference in velocity estimates between transcranial Doppler (TCD) systems when measuring high velocities (∼200 cm/s) close to the threshold for sickle cell disease stroke prevention. As we were unable to identify a suitable commercial TCD phantom, a middle cerebral artery (MCA) flow phantom was developed to evaluate velocity estimates from different devices under controlled conditions. Time-averaged velocity estimates were obtained using two TCD devices: a Spencer Technologies ST3 Doppler system (ST3 PMD150, Spencer Technologies, Seattle, WA, USA) and a DWL Dopplerbox (DWL Compumedics, SN-300947, Singen, Germany). These were compared with velocity estimates obtained using a Zonare duplex scanner (Zonare Medical Systems, Mountain View, CA, USA), with timed collection of fluid as the gold standard. Bland-Altman analysis was performed to compare measurements between devices. Our tests confirmed that velocities measured with the DWL TCD system were +4.1 cm/s (+3.7%; limits of agreement [LoA]: 2%, 5%; p = 0.03) higher than the Spencer system when measuring a velocity 110 cm/s and +12 cm/s higher (+5.7 %; LoA: 4.8%, 6.6%; p = 0.03) when measuring velocities of 210 cm/s, close to the diagnostic threshold for stroke intervention. We found our MCA phantom to be a valuable tool for systematically quantifying differences in TCD velocity estimates between devices, confirming that the DWL system gave consistently higher readings than the Spencer ST3 system. Differences become more pronounced at high velocities, which explains why they were not identified earlier. Our findings have clinical implications for centers using TCD to monitor patients with sickle cell disease, as extra care may be needed to adjust for bias between manufacturers when making treatment decisions about children with sickle cell with velocities close to the diagnostic threshold.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Alablani, Janus, Pallett, et al. Development of a Flow Phantom for Transcranial Doppler Ultrasound Quality Assurance. Ultrasound Med Biol. 2022. PMID: 36038392. doi:10.1016/j.ultrasmedbio.2022.07.002

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。