新生儿猪肺超声诱导肺毛细血管出血研究

Lung Ultrasound Induction of Pulmonary Capillary Hemorrhage in Neonatal Swine.
PubMed 文献 · Ultrasound Med Biol · 2022-08-25

本研究用新生雌性仔猪在温浴中经肋间行肺超声,比较三种设备与探头及剪切波弹性成像对肺毛细血管出血的诱导作用;C1-5探头在50%与100%声输出下出血阈值为0.53 MIIS,L4-12t未诱导出血,7L探头诱导扫描平面出血,自动B线工具对C1-5所致出血未计数B线但换用L4-12t后可计数,SWE在推脉冲位置诱导出血并随作用时间累积,研究支持新生儿肺超声使用MI<0.4的安全建议。

要点

  • 新生儿肺超声应控制机械指数,建议MI<0.4以降低肺毛细血管出血风险。
  • 不同探头诱导出血能力不同,L4-12t未诱导出血而C1-5与7L可诱导。
  • 自动B线工具可能漏检C1-5探头所致出血,换用L4-12t探头可计数相关B线。

核心结论

  • 肺超声可在新生仔猪中诱导肺毛细血管出血,支持新生儿肺超声使用MI<0.4的安全建议。

研究设计与样本

  • 研究为新生雌性仔猪实验,麻醉后垂直固定于温浴中,在2或3个肋间进行扫描。
  • 使用GE Venue R1床旁系统配C1-5与L4-12t探头、GE Vivid 7 Dimension配7L探头、SuperSonic Imagine配SL15-4探头及剪切波弹性成像。
  • 以水听器测量计算原位机械指数MIIS,取充气肺组织评分肺毛细血管出血面积作为参考标准。

关键影像表现/征象

  • C1-5探头在50%与100%声输出下可诱导肺毛细血管出血,线性回归得到出血阈值0.53 MIIS(r2=0.42)。
  • L4-12t探头未诱导肺毛细血管出血,7L探头在扫描平面诱导出血区。
  • C1-5探头所致出血未被Venue R1自动B线工具计为B线,但后续用L4-12t探头自动工具扫描时可计数与出血相关的B线。
  • 剪切波弹性成像在推脉冲位置诱导出血,3、30与300秒作用后出血累积,出血面积以0.33 mm2/s累积并指数上升至最大18.4 mm2(r2=0.61)。

诊断标准与关键数值

  • C1-5探头在50%与100%声输出条件下,肺毛细血管出血阈值为0.53 MIIS(线性回归,r2=0.42)。
  • 剪切波弹性成像诱导出血面积随作用时间累积,累积速率0.33 mm2/s,指数上升最大18.4 mm2(r2=0.61)。

临床应用与阅片要点

  • 新生儿肺超声检查时应关注机械指数,尽量保持MI<0.4以减少肺毛细血管出血风险。
  • 使用自动B线工具时需注意探头相关差异,C1-5探头所致出血可能不被计数,必要时可换用L4-12t探头复核。
  • 剪切波弹性成像在新生儿肺应用时需注意推脉冲可能诱导出血,应限制作用时间与输出。

局限性

  • 研究为新生仔猪动物实验,结果外推至人类新生儿需谨慎。
  • 原文未报告样本量、中心数及前瞻性或回顾性设计等具体信息。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

This study investigated induction of pulmonary capillary hemorrhage (PCH) in neonatal pigs (piglets) using three different machines: a GE Venue R1 point-of-care system with C1-5 and L4-12t probes, a GE Vivid 7 Dimension with a 7L probe and a SuperSonic Imagine machine with an SL15-4 probe and shear wave elastography (SWE). Female piglets were anesthetized, and each was mounted vertically in a warm bath for scanning at two or three intercostal spaces. After aiming at an innocuous output, the power was raised for a test exposure. Hydrophone measurements were used to calculate in situ values of mechanical index (MIIS). Inflated lungs were removed and scored for PCH area. For the C1-5 probe at 50% and 100% acoustical output (AO), a PCH threshold of 0.53 MIIS was obtained by linear regression (r2 = 0.42). The L4-12t probe did not induce PCH, but the 7L probe induced zones of PCH in the scan planes. The Venue R1 automated B-line tool applied with the C1-5 probe did not detect PCH induced by the C1-5 probe as B-line counts. However, when PCH induced by C1-5 and 7L exposures were subsequently scanned with the L4-12t probe using the automated tool, B-lines were counted in association with the PCH. The SWE induced PCH at push-pulse positions for 3, 30 and 300 s of SWE with PCH accumulating at 0.33 mm2/s and an exponential rise to a maximum of 18.4 mm2 (r2 = 0.61). This study demonstrated the induction of PCH by LUS of piglets, and supports the safety recommendation for use of MIs <0.4 in neonatal LUS.