不同厂商二维与点剪切波弹性成像对慢性肝病肝纤维化分期的前瞻性对比研究
单中心前瞻性研究纳入121例慢性肝病患者,在肝活检前进行pSWE和2-D SWE测量,以Ishak评分为参考标准。两种SWE均与纤维化分期显著相关,2-D SWE的AUC略高于pSWE,技术失败率均为零。研究表明2-D SWE可准确评估肝纤维化,性能与pSWE相当。
要点
- 2-D SWE与pSWE均与肝纤维化分期显著相关,无技术失败。
- 2-D SWE鉴别F≥3的AUC为0.884,鉴别F≥5的AUC为0.926。
- pSWE鉴别F≥3的AUC为0.855,鉴别F≥5的AUC为0.890。
核心结论
- 2-D SWE对肝纤维化评估可行且准确,与两台不同厂商pSWE性能相当。
研究设计与样本
- 单中心前瞻性研究,121例成人慢性肝病患者,年龄18-70岁(中位45岁),肝活检前即刻行pSWE和2-D SWE测量。
- 以肝组织学Ishak评分为参考标准,使用Kruskal-Wallis检验、Spearman相关和ROC曲线分析。
诊断标准与关键数值
- pSWE与2-D SWE的SWV平均差值为0.0223,一致性界限为-1.1009至1.1145。
- pSWE与2-D SWE的SWV均与纤维化分期显著相关,Spearman ρ分别为0.606(p<0.0001)和0.722(p<0.001)。
- 鉴别F≥3的AUC:pSWE为0.855(95%CI 0.778-0.932),2-D SWE为0.884(95%CI 0.817-0.951)。
- 鉴别F≥5的AUC:pSWE为0.890(95%CI 0.826-0.954),2-D SWE为0.926(95%CI 0.88-0.973)。
- 采用Youden指数确定最佳截断点。
临床应用与阅片要点
- 2-D SWE可作为肝纤维化无创评估工具,尤其适用于需要二维定量分析的场景。
- pSWE与2-D SWE测值一致性良好,但2-D SWE相关性略高,报告时可注明所用技术。
局限性
- 单中心前瞻性研究,样本量121例,结论外推需多中心验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The purpose of this study was to assess the utility of 2-D shear wave elastography (SWE) in assessing liver fibrosis in patients with chronic liver disease by comparing its performance with that of point shear wave elastography (pSWE) using liver histological staging as the reference standard. In this ethics committee-approved, single-institution prospective study, pSWE and 2-D SWE velocity measurements were obtained in 121 adult patients (age: 18-70 y, median: 45 y) immediately before a liver biopsy for chronic liver disease. Shear wave velocity (SWV) and Ishak scores were compared using the Kruskal-Wallis test, Spearman's correlation and receiver operating characteristic (ROC) curve analysis. Youden's index was used to determine the optimal cutoff point. There was no technical failure using pSWE and 2-D SWE. The mean difference for SWV between pSWE and 2-D SWE was 0.0223 (limits of agreement: -1.1009, 1.1145). Values for both pSWE and 2-D SWE were significantly correlated with fibrosis stage (Spearman's ρ = 0.606, p < 0.0001; ρ = 0.722, p < 0.001 respectively). The area under the ROC curve differentiating F ≥3 was 0.855 (95% confidence interval: 0.778-0.932) for pSWE and 0.884 (95% CI: 0.817-0.951) for 2-D SWE. The AUC for differentiating F ≥5 was 0.890 (95% CI: 0.826-0.954) for pSWE and 0.926 (95% CI: 0.88-0.973) for 2-D SWE. This study indicates that 2-D SWE provides feasible and accurate assessment of liver fibrosis, comparable to that provided by pSWE from two different manufacturers' machines.
引用
- Fang, Rafailidis, Konstantatou, et al. Comparison Between Different Manufacturers' 2-D and Point Shear Wave Elastography Techniques in Staging Liver Fibrosis in Chronic Liver Disease Using Liver Biopsy as the Reference Standard: A Prospective Study. Ultrasound Med Biol. 2022. PMID: 36008215. doi:10.1016/j.ultrasmedbio.2022.06.009
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。