超快平面波与高帧频聚焦发射在体估计脉搏波速度的比较
本研究在14名健康志愿者中连续采集颈动脉超声射频数据,比较快速聚焦成像(FFI)与平面波成像(PWI)在体估计局部脉搏波速度(PWV)的准确性。两种方法估计的PWV无统计学显著差异,重搏切迹(DN)时间参考点的PWV和R²平均高于收缩期足(SF),40–80 Hz截止频率提供最准确PWV,两种方法临床适用性相当但作者偏好DN处PWV。
要点
- FFI与PWI估计的颈动脉局部PWV无显著差异,两者临床适用性相当。
- DN时间参考点的PWV和R²平均高于SF,作者更倾向采用DN处PWV。
- 低通滤波截止频率40–80 Hz时PWV估计最准确。
英文原文(摘要)· 点击展开
Ultrasound-based local pulse wave velocity (PWV) estimation, as a measure of arterial stiffness, can be based on fast focused imaging (FFI) or plane wave imaging (PWI). This study was aimed at comparing the accuracy of in vivo PWV estimation using FFI and PWI. Ultrasound radiofrequency data of carotid arteries were acquired in 14 healthy volunteers (25-57 y) by executing the FFI (12 lines, 7200 Hz) and PWI (128 lines, 2000 Hz) methods consecutively. PWV was derived at two time-reference points, dicrotic notch (DN) and systolic foot (SF), for multiple pressure cycles by fitting a linear function through the positions of the peaks of low-pass filtered wall acceleration curves as a function of time. The accuracy of PWV estimation was determined for various cutoff frequencies (10-200 Hz). No statistically significant difference was observed between PWVs estimated by both approaches. The PWV and R2 at DN were higher, on average, than those at SF (PWV/R2: FFI SF 5.5/0.92, FFI DN 6.1/0.92; PWI SF 5.4/0.89, PWI DN 6.3/0.95). The use of cutoff frequencies between 40 and 80 Hz provided the most accurate PWVs. Both methods seemed equally suitable for use in clinical practice, although we have a preference for the PWV at DN given the higher R2 values.
引用
- van den Bos-van de Steeg, Fekkes, Saris, et al. In Vivo Comparison of Pulse Wave Velocity Estimation Based on Ultrafast Plane Wave Imaging and High-Frame-Rate Focused Transmissions. Ultrasound Med Biol. 2022. PMID: 35999121. doi:10.1016/j.ultrasmedbio.2022.06.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。