文献速递心血管外周血管MR

机器学习自动评估4D flow CMR主动脉旋转血流与壁面剪切应力

Machine learning for the automatic assessment of aortic rotational flow and wall shear stress from 4D flow cardiac magnetic resonance imaging.
PubMed 文献 · Eur Radiol · 2022-08-17

自动分割Dice评分中位数0.949,SFRR与手动定量ICC>0.9,IRF及轴向/周向WSS ICC>0.75,表明全自动分析复杂主动脉血流动力学可行。

要点

  • 自动分割主动脉Dice评分中位数0.949,与手动分割一致性高。
  • SFRR自动与手动定量ICC>0.9,IRF及轴向/周向WSS ICC>0.75。
  • 肺动脉分叉处标志检测精度低于专家,可能影响分析平面定位。

诊断标准与关键数值

  • 自动主动脉分割中位Dice评分0.949,平均对称表面距离0.839(0.632-1.071)mm。
  • SFRR自动与手动定量ICC>0.9,IRF及轴向和周向WSS的ICC>0.75。
  • 主动脉标志定位精度在窦管交界及第一、第三上主动脉血管与专家相当(p=0.513、0.592、0.905),肺动脉分叉处较低(p=0.028)。

局限性

  • 单中心前瞻性研究,外部泛化性未验证。
  • 标志检测在肺动脉分叉处精度低于专家,可能影响该区域分析平面定位。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: Three-dimensional (3D) time-resolved phase-contrast cardiac magnetic resonance (4D flow CMR) allows for unparalleled quantification of blood velocity. Despite established potential in aortic diseases, the analysis is time-consuming and requires expert knowledge, hindering clinical application. The present research aimed to develop and test a fully automatic machine learning-based pipeline for aortic 4D flow CMR analysis.

METHODS: Four hundred and four subjects were prospectively included. Ground-truth to train the algorithms was generated by experts. The cohort was divided into training (323 patients) and testing (81) sets and used to train and test a 3D nnU-Net for segmentation and a Deep Q-Network algorithm for landmark detection. In-plane (IRF) and through-plane (SFRR) rotational flow descriptors and axial and circumferential wall shear stress (WSS) were computed at ten planes covering the ascending aorta and arch.

RESULTS: Automatic aortic segmentation resulted in a median Dice score (DS) of 0.949 and average symmetric surface distance of 0.839 (0.632-1.071) mm, comparable with the state of the art. Aortic landmarks were located with a precision comparable with experts in the sinotubular junction and first and third supra-aortic vessels (p = 0.513, 0.592 and 0.905, respectively) but with lower precision in the pulmonary bifurcation (p = 0.028), resulting in precise localisation of analysis planes. Automatic flow assessment showed excellent (ICC > 0.9) agreement with manual quantification of SFRR and good-to-excellent agreement (ICC > 0.75) in the measurement of IRF and axial and circumferential WSS.

CONCLUSION: Fully automatic analysis of complex aortic flow dynamics from 4D flow CMR is feasible. Its implementation could foster the clinical use of 4D flow CMR.

KEY POINTS: • 4D flow CMR allows for unparalleled aortic blood flow analysis but requires aortic segmentation and anatomical landmark identification, which are time-consuming, limiting 4D flow CMR widespread use. • A fully automatic machine learning pipeline for aortic 4D flow CMR analysis was trained with data of 323 patients and tested in 81 patients, ensuring a balanced distribution of aneurysm aetiologies. • Automatic assessment of complex flow characteristics such as rotational flow and wall shear stress showed good-to-excellent agreement with manual quantification.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Garrido-Oliver, Aviles, Córdova, et al. Machine learning for the automatic assessment of aortic rotational flow and wall shear stress from 4D flow cardiac magnetic resonance imaging. Eur Radiol. 2022. PMID: 35976395. doi:10.1007/s00330-022-09068-9

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。