无龙门架锥形束CT联合迭代金属伪影减少用于四肢骨术后随访影像评估

Combining gantry-free cone-beam computed tomography with iterative metal artefact reduction for surgical follow-up imaging of the appendicular skeleton.
PubMed 文献 · Eur J Radiol · 2022-08-10

结果显示CBCT对螺钉脱位、植入物松动、骨折块移位及延迟愈合的诊断准确率均显著高于X线,术后并发症检测的敏感度和特异度至少为95.8%和98.1%,而X线仅为33.3%和92.86%。

要点

  • 四肢骨术后评估中,CBCT对螺钉脱位、植入物松动、骨折块移位及延迟愈合的检出准确率均优于X线。
  • CBCT检测术后并发症的敏感度至少95.8%、特异度98.1%,明显高于X线的33.3%和92.86%。
  • CBCT中位剂量面积乘积为27.98 dGy*cm2,X线为0.2 dGy*cm2,需权衡辐射剂量。

局限性

  • 回顾性研究设计。
  • 单中心研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

PURPOSE: Post-surgical evaluation of osteosynthesis material and adjacent tissue can be challenging in both radiography and cross-sectional imaging. This study investigates the performance of a multi-purpose X-ray scanner with cone-beam CT (CBCT) function and iterative metal artefact reduction capabilities in patients after osteoplasty of the appendicular skeleton.

METHOD: Eighty individuals who underwent both conventional X-ray imaging and CBCT after osteoplasty of the hand/wrist (48), elbow (14), or ankle/foot (18) with the gantry-free twin robotic system were retrospectively enrolled. Radiological reports from clinical routine for both imaging modalities were retrospectively analyzed and compared with consensus expert reading by two musculoskeletal specialists serving as the standard of reference. Findings of screw dislocation or implant loosening, fragment displacement, and delayed healing were compared between X-ray and CBCT reports using the McNemar test.

RESULTS: The median dose-area-product of CBCT and X-ray scans amounted to 27.98 and 0.2 dGy*cm2, respectively. Diagnostic accuracy for screw dislocation was superior in CBCT compared to standard radiograms (98.8 % vs 83.8 %; p = 0.002). Implant loosening (98.8 % vs 86.3 %; p = 0.006), fragment displacement (98.8 % vs 85.0 %; p < 0.001), and delayed healing (97.5 % vs 88.8 %; p = 0.016) were also more reliably detected in CBCT. Employing CBCT, postoperative complications were detected with a sensitivity and specificity of at least 95.8 % and 98.1 %, compared to 33.3 % and 92.86 % in radiography.

CONCLUSIONS: With superior accuracy for various osteoplasty-related complications, the CBCT scan mode of a gantry-free twin robotic X-ray system with iterative metal artefact reduction aids post-surgical assessment in the appendicular skeleton.

来源:Eur J Radiol,查看原文。版权归原作者所有。

引用

  1. Patzer, Grunz, Huflage, et al. Combining gantry-free cone-beam computed tomography with iterative metal artefact reduction for surgical follow-up imaging of the appendicular skeleton. Eur J Radiol. 2022. PMID: 35973302. doi:10.1016/j.ejrad.2022.110465

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。