锡滤过100 kV超低剂量腹部CT检测尿路结石:510例对比研究

Tin-filtered 100 kV Ultra-low-dose Abdominal CT for Calculi Detection in the Urinary Tract: A Comparative Study of 510 Cases.
PubMed 文献 · Acad Radiol · 2022-08-11

该回顾性单中心研究对比了锡预滤过100 kVp超低剂量方案与常规低剂量方案在尿路结石检测中的表现,共纳入510例。结果显示两组结石检出频率无显著差异,超低剂量组有效剂量较常规低剂量组降低62.35%,图像噪声无显著差异。

要点

  • 锡滤过100 kVp超低剂量CT可检出尿路结石且不降低检出频率。
  • 超低剂量方案有效剂量较常规低剂量降低62.35%。
  • 两组图像噪声无显著差异,图像质量相当。

核心结论

  • 锡预滤过100 kVp超低剂量扫描可在不损害图像质量的前提下,将尿路结石检测的有效剂量较常规低剂量CT降低达62%。

研究设计与样本

  • 回顾性单中心研究,共纳入510例,低剂量组290例(女96例,年龄49±16岁,BMI 27.23±5.60 kg/m2),超低剂量组220例(女84例,年龄47±17岁,BMI 26.82±5.62 kg/m2),两组年龄(p=0.132)和BMI(p=0.207)无显著差异。
  • 超低剂量组采用锡预滤过、标准管电压100 kVp;低剂量组为临床已建立的专用低剂量方案;以图像噪声进行定量评估。

诊断标准与关键数值

  • 两组结石检出频率无显著差异(p=0.596)。
  • 超低剂量组有效剂量1.16 mSv(IQR 0.89-1.54 mSv),较低剂量组3.08 mSv(IQR 2.22-4.02 mSv)降低62.35%。
  • 图像噪声低剂量组为18.90(IQR 17.39-21.20),超低剂量组为18.69(IQR 17.30-21.62),两组无显著差异(p=0.793)。

临床应用与阅片要点

  • 在尿路结石检出中可考虑采用锡滤过100 kVp超低剂量腹部CT方案,以显著降低辐射剂量。
  • 图像噪声在两组间无显著差异,提示超低剂量方案未明显影响图像质量。

局限性

  • 本研究为回顾性单中心设计,可能存在选择偏倚。
  • 原文未报告敏感度、特异度、AUC等诊断准确性指标,无法评估诊断效能。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: For detection of urinary calculi, unenhanced low-dose computed tomography is the method of choice, outperforming radiography and ultrasound. This retrospective monocentric study aims to compare a clinically established, dedicated low-dose imaging protocol for detection of urinary calculi with an ultra-low-dose protocol employing tin prefiltration at a standardized tube voltage of 100 kVp.

METHODS: Two study arms included a total of 510 cases. The "low-dose group" was comprised of 290 individuals (96 women; age 49 ± 16 years; BMI 27.23 ± 5.60 kg/m2). The "ultra-low-dose group" with Sn100 kVp consisted of 220 patients (84 women; age 47 ± 17 years; BMI 26.82 ± 5.62 kg/m2). No significant difference was ascertained for comparison of age (p = 0.132) and BMI (p = 0.207) between cohorts. For quantitative assessment of image quality, image noise was assessed.

RESULTS: No significant difference regarding frequency of calculi detection was found between groups (p = 0.596). Compared to the low-dose protocol (3.08 mSv; IQR 2.22-4.02 mSv), effective dose was reduced by 62.35% with the ultra-low-dose protocol employing spectral shaping (1.16 mSv; IQR 0.89-1.54 mSv). Image noise was calculated at 18.90 (IQR 17.39-21.20) for the low-dose protocol and at 18.69 (IQR 17.30-21.62) for the ultra-low-dose spectral shaping protocol. No significant difference was ascertained for comparison between groups (p = 0.793).

CONCLUSION: For urinary calculi detection, ultra-low-dose scans utilizing spectral shaping by means of tin prefiltration at 100 kVp allow for considerable dose reduction of up to 62% over conventional low-dose CT without compromising image quality.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Kunz, Grunz, Halt, et al. Tin-filtered 100 kV Ultra-low-dose Abdominal CT for Calculi Detection in the Urinary Tract: A Comparative Study of 510 Cases. Acad Radiol. 2022. PMID: 35963837. doi:10.1016/j.acra.2022.07.013

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。