用于二维脑成像的全波形反演低频超声采集装置的设计与构建
本研究设计并构建了一种用于经颅脑成像的低频环形超声换能器阵列,采用全波形反演进行图像重建。装置使用40个换能器组成环形阵列,通过旋转实现高密度覆盖并获取冗余信号。实验室结果显示该装置能对模拟脑和颅骨的2.5维体模进行全波形反演成像,以毫米级分辨率重建解剖和声学特性,据作者所知是首个用于类经颅成像的原型。
要点
- 全波形反演可克服传统超声因颅骨畸变无法脑成像的限制。
- 40换能器环形阵列结合旋转可实现高密度覆盖和冗余信号采集。
- 该原型成功对2.5维脑-颅骨模拟体模进行全波形反演成像。
核心结论
- 作者构建了40换能器环形低频超声阵列原型,实验室结果表明其能利用全波形反演对2.5维脑-颅骨模拟体模成像,并以毫米级分辨率重建解剖和声学特性,据作者所知是首个用于类经颅成像的原型。
研究设计与样本
- 研究为设备设计与构建研究,探究环形超声换能器阵列的不同设计以优化经颅全波形反演成像所需的阵元数和旋转次数,并构建40换能器环形阵列原型进行实验室体模验证。
诊断标准与关键数值
- 装置使用小直径低频换能器,超声可穿过颅骨传播并具有良好信噪比。
- 采用旋转实现目标高密度覆盖并采集有利于全波形反演的冗余信号。
- 40换能器环形阵列可对2.5维脑-颅骨模拟体模进行全波形反演成像,以毫米级分辨率重建解剖和声学特性。
局限性
- 目前尚无用于经颅应用的全波形反演原型。
- 实验室结果仅基于2.5维脑-颅骨模拟体模,尚未在临床中验证。
- 作者在讨论中提出这些发现对设计可能临床应用的3维重建系统的意义,但未报告临床数据。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
The main techniques used to image the brain and obtain structural data are magnetic resonance imaging and X-ray computed tomography. These techniques produce images with high spatial resolution, but with the disadvantage of requiring very large equipment with special installation needs. In addition, X-ray tomography uses ionizing radiation, which limits their use. Ultrasound imaging is a safe technology that is delivered using compact and mobile devices. However, conventional ultrasound reconstruction techniques have failed to obtain images of the brain because of, fundamentally, the presence of the skull and the distortion that it produces on ultrasound. Recent studies have indicated that full-waveform inversion, a computational technique originally from Earth science, has the potential to generate accurate 3-D images of the brain. This technology can overcome the limitations of conventional ultrasound imaging, but a prototype for transcranial applications does not yet exist. Here, we investigate different designs of an annular array of ultrasound transducers to optimize the number of elements and rotations needed to conduct transcranial imaging with full-waveform inversion. This device uses small-diameter, low-frequency transducers that readily propagate ultrasound through the skull with good signal-to-noise ratios. It also incorporates the use of rotations to produce a high-density coverage of the target and acquire redundant traces that are beneficial for full-waveform inversion. We have built a ring of 40 transducers to illustrate that this design is capable of reconstructing images of the brain, retrieving its anatomy and acoustic properties with millimeter resolution. Laboratory results reveal the ability of this device to successfully image a 2.5-D brain- and skull-mimicking phantom using full-waveform inversion. To our knowledge, this is the first prototype ever used for transcranial-like imaging. The importance of these findings and their implications for the design of a 3-D reconstruction system with possible clinical applications are discussed.
引用
- Cudeiro-Blanco, Cueto, Bates, et al. Design and Construction of a Low-Frequency Ultrasound Acquisition Device for 2-D Brain Imaging Using Full-Waveform Inversion. Ultrasound Med Biol. 2022. PMID: 35902276. doi:10.1016/j.ultrasmedbio.2022.05.023
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。