迭代去噪三维SPACE联合内体积激发与同时多层BLADE在1.5T女性盆腔磁共振成像优化中的定量评估

Quantitative Assessment of Iteratively Denoised 3D SPACE with Inner-Volume Excitation and Simultaneous Multi-Slice BLADE for Optimizing Female Pelvis Magnetic Resonance Imaging at 1.5 T.
PubMed 文献 · Acad Radiol · 2022-07-20

前瞻性纳入38例女性,比较标准BLADE、同时多层BLADE(提高时间分辨率或空间分辨率)及迭代去噪三维SPACE的图像质量;SMS BLADE(sr)因整体图像印象、器官勾画和病灶分界略优而被选为首选序列,SMS BLADE(tr)可在不损害图像质量的前提下缩短采集时间,三维SPACE图像质量较低但可减少检查时间并通过回顾性多平面重建改善工作流程。

要点

  • SMS BLADE(sr)在整体图像印象、器官勾画和病灶分界上略优于标准BLADE,但差异无统计学意义,被选为首选序列。
  • SMS BLADE(tr)与标准BLADE图像质量相似,运动伪影显著更少(p<0.01),采集时间从2分59秒缩短至1分20秒。
  • 三维SPACE整体图像印象和运动伪影评分显著低于标准BLADE(p=0.01和p<0.01),但可回顾性多平面重建以缩短检查时间。

核心结论

  • SMS BLADE(sr)是女性盆腔MRI的首选序列,空间分辨率高于标准BLADE;SMS BLADE(tr)也可在不损害图像质量的情况下缩短采集时间;三维SPACE图像质量较低,但可减少检查时间并通过回顾性多平面重建改善工作流程。

研究设计与样本

  • 前瞻性研究纳入38例因临床指征行子宫MRI的女性(平均年龄44±15岁),比较标准BLADE(采集时间2分59秒,层厚4mm)、SMS BLADE(tr)(层厚4mm,采集时间1分20秒)、SMS BLADE(sr)(层厚3mm,采集时间3分16秒)及矢状位三维SPACE(采集时间5分36秒),并回顾性重建宫颈短轴和子宫体长轴1mm与3mm层厚图像。

诊断标准与关键数值

  • 采用5点Likert量表评估整体图像印象、解剖/器官勾画、病灶分界和运动伪影;标准BLADE、SMS BLADE(sr)和SMS BLADE(tr)的整体图像印象相似,三维SPACE显著低于标准BLADE(p=0.01);SMS BLADE(tr)和(sr)的运动伪影均显著少于标准BLADE(p<0.01和p=0.01),两者间无显著差异,三维SPACE显著低于标准BLADE(p<0.01);图像质量在所有序列和所有病例中均被评为最不具诊断价值的标准。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: High-resolution T2-weighted magnetic resonance imaging (MRI) of the pelvis is the main technique used for diagnosing benign and malignant uterine diseases. However, the procedure may be time-consuming and requires training and experience. Therefore, this study was performed to compare the image quality of standard clinical BLADE (stBLADE) with a prototypical accelerated simultaneous multi-slice (SMS) BLADE procedure with either improved temporal resolution (tr) at the same slice thickness (SL) or improved spatial resolution (sr) with the same examination time and a prototypical isotropic 3D SPACE procedure with inner-volume excitation and iterative denoising.

MATERIALS AND METHODS: Patients who underwent clinically indicated MRI of the uterus were included in this prospective study and underwent stBLADE (acquisition time, 2 min 59 s; SL, 4 mm) and SMS BLADE (tr) with the same SL (4 mm) but reduced examination time (1 min 20 s) as well as SMS BLADE (sr) with thinner slices (3 mm) and comparable examination time (3 min 16 s). In addition, 3D SPACE was acquired in a sagittal orientation (5 min 36 s). The short axis of the cervix and the long axis of the corpus uteri were reconstructed in 1-mm and 3-mm SLs, retrospectively. Subjective overall image impression, delineation of anatomy/organs, lesion demarcation, and motion artifacts were assessed using a 5-point Likert scale and compared among the different techniques. The preferred sequence was then selected by three independent assessors.

RESULTS: The analysis was based on 38 women (mean age, 44 ± 15 years). The overall image impression was similar for stBLADE, SMS BLADE (sr), and SMS BLADE (tr) but was significantly lower for 3D SPACE than stBLADE (p = 0.01). SMS BLADE (sr) was considered the preferred sequence because of slightly better performance in terms of overall image impression, organ delineation, and lesion demarcation, but without statistical significance. Both SMS BLADE (tr) and (sr) produced significantly fewer motion artifacts than stBLADE (p < 0.01 and p = 0.01), with no significant difference between SMS BLADE (tr) and (sr), while 3D SPACE had a significantly lower rating than stBLADE (p < 0.01). Image quality was rated as the least diagnostic criterion in all sequences and all cases.

CONCLUSION: SMS BLADE (sr) was the preferred sequence for MRI of the female pelvis, with higher sr than stBLADE. SMS BLADE (tr) may also be used to reduce the acquisition time without compromising image quality. Despite its lower image quality, 3D SPACE can also reduce the examination time and improve the workflow because of the possibility of retrospective multiplanar reconstructions.

来源:Acad Radiol,查看原文。版权归原作者所有。