以患者为中心的自助压迫乳腺X线摄影在临床实践中的随机试验:与标准压迫的比较

Patient-centred care with self-compression mammography in clinical practice: a randomized trial compared to standard compression.
PubMed 文献 · Eur Radiol · 2022-07-22

一项随机试验纳入495名接受乳腺X线摄影的女性,比较自助压迫与放射技师压迫对腺体剂量、图像质量和随访依从性的影响。结果显示自助压迫达到更强压迫力且图像质量相似,但腺体剂量几乎无变化(1.90 vs 1.93 mGy),随访参与率亦无差异。

要点

  • 自助压迫与技师压迫的图像质量相似(放射科医生判断p=0.90,技师判断p=0.32)。
  • 自助压迫达到更强压迫力(114.5 vs 10.25 daN,p<.001),但腺体剂量几乎无影响(1.90 vs 1.93 mGy,p=.47)。
  • 两种压迫方式对后续乳腺X线摄影参与率无显著影响(p=0.47)。

核心结论

  • 自助压迫达到更强乳房压迫且图像质量相当,但未显著降低腺体剂量,随访参与率两组亦相似。

研究设计与样本

  • 随机对照试验,纳入495名女性(平均年龄57岁±14),自助压迫组245例,放射技师压迫组250例。

诊断标准与关键数值

  • 图像质量由两名独立放射科医生和两名放射技师盲法评估,两组相似(放射科医生判断p=0.90;技师判断p=0.32)。
  • 自助压迫组压迫力114.5 daN vs 技师压迫组10.25 daN(p<.001)。
  • 自助压迫组腺体剂量1.90 mGy vs 技师压迫组1.93 mGy(p=.47)。
  • 中度/重度不适报告率自助压迫组7.8% vs 技师压迫组9.6%(p=.77)。
  • 中位疼痛评分两组均为4.0(p=.55)。

临床应用与阅片要点

  • 自助压迫可达到更强压迫力且图像质量与技师压迫相似,但未降低腺体剂量,对后续乳腺X线摄影参与率无影响。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVE: To test the efficacy of self- compared to radiographer-led compression to reduce the average glandular dose without affecting image quality and compliance to follow-up mammography.

MATERIALS AND METHODS: Women presenting for mammography for breast cancer follow-up, symptoms, opportunistic screening, or familial risk were asked to participate and, if willing, were randomized to self-compression or radiographer-led compression. Image quality was assessed blindly by two independent radiologists and two radiographers. Pain and discomfort were measured immediately after mammography and their recall was asked when the women participated in the follow-up mammogram, 1 or 2 years later.

RESULTS: In total, 495 women (mean age 57 years +/-14) were enrolled, 245 in the self-compression and 250 radiographer-compression arms. Image quality was similar in the two arms (radiologists' judgement p = 0.90; radiographers' judgement p = 0.32). A stronger compression force was reached in the self- than in the radiographer-arm (114.5 vs. 10.25 daN, p < .001), with a 1.7-mm reduction in thickness (p = .14), and almost no impact on dose per exam (1.90 vs. 1.93 mGy, p = .47). Moderate/severe discomfort was reported by 7.8% vs 9.6% (p = .77) and median pain score was 4.0 in both arms (p = .55). Median execution time was 1 min longer with self-compression (10.0 vs. 9.1 min, p < 0.001). No effect on subsequent mammography was detectable (p = 0.47).

CONCLUSION: Self-compression achieved stronger compression of the breast, with comparable image quality, but did not substantially reduce glandular dose. The proportion of women who attended follow-up mammography was also similar in the two groups.

TRIAL REGISTRATION: clinicaltrials.gov NCT04009278 KEY POINTS: • In mammography, appropriate compression is essential to obtain high image quality and reduce dose. Compression causes pain and discomfort. • Self-compression has been proposed to reach better compression and possibly increase participation in mammography. • In a randomized trial, self-compression reached stronger compression of the breast, with comparable image quality but with no glandular dose reduction or impact on participation in follow-up mammography.

来源:Eur Radiol,查看原文。版权归原作者所有。

引用

  1. Iotti, Giorgi Rossi, Canovi, et al. Patient-centred care with self-compression mammography in clinical practice: a randomized trial compared to standard compression. Eur Radiol. 2022. PMID: 35869315. doi:10.1007/s00330-022-09002-z

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。