文献速递乳腺技术与安全PET

体内切伦科夫成像放疗事件检测中定量指标的性能比较

Performance comparison of quantitative metrics for analysis of in vivo Cherenkov imaging incident detection during radiotherapy.
PubMed 文献 · Br J Radiol · 2022-07-27

研究用模拟乳腺放疗的血管化体模,系统平移并采集切伦科夫图像,比较互信息与γ通过率对位置误差的响应,并与Dice系数和平均一致性距离对比。互信息和γ通过率均能单调检测各方向平移,γ通过率对强度变化最敏感,但两者均不能检测射束形状失配。

要点

  • 互信息和γ通过率可单调检测各方向平移误差,随平移增大而恶化。
  • γ通过率对强度变化(如红斑或 bolus 组织空气间隙)检测最敏感。
  • 互信息和γ通过率均不能检测射束形状失配,Dice 和 MDC 可检测。

核心结论

  • 不同放疗事件可通过比较分次内和分次间切伦科夫图像并用相应的图像相似性指标来检测,指标选择随事件类型而异。

研究设计与样本

  • 体模研究:使用模拟患者血管(切伦科夫图像中的生物标志)的拟人体模进行乳腺放疗模拟,系统平移各方向并采集切伦科夫图像,比较互信息、γ通过率与 Dice 系数、平均一致性距离(MDC)的响应,并分析含其他事件的患者图像以验证最佳检测算法。

诊断标准与关键数值

  • 互信息和γ通过率均能登记各方向位置平移,且随平移增大单调恶化。
  • 强度变化(可能源于红斑或 bolus 组织空气间隙)最易被γ通过率检测。
  • 射束形状失配(如剂量照射到非预期区域)不能被互信息和γ通过率检测,而 Dice 和 MDC 可检测。

临床应用与阅片要点

  • 比较分次内和分次间切伦科夫图像时,应根据事件类型选择相应的图像相似性指标。
  • 未来工作将确定各指标的适当阈值以实现自动标记。
  • 分类不同算法用于检测各种放疗事件,有助于开发自动标记系统,消除人工审阅切伦科夫图像的负担。

局限性

  • 未来工作将确定各指标的适当阈值以实现自动标记。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Examine the responses of multiple image similarity metrics to detect patient positioning errors in radiotherapy observed through Cherenkov imaging, which may be used to optimize automated incident detection.

METHODS: An anthropomorphic phantom mimicking patient vasculature, a biological marker seen in Cherenkov images, was simulated for a breast radiotherapy treatment. The phantom was systematically shifted in each translational direction, and Cherenkov images were captured during treatment delivery at each step. The responses of mutual information (MI) and the γ passing rate (%GP) were compared to that of existing field-shape matching image metrics, the Dice coefficient, and mean distance to conformity (MDC). Patient images containing other incidents were analyzed to verify the best detection algorithm for different incident types.

RESULTS: Positional shifts in all directions were registered by both MI and %GP, degrading monotonically as the shifts increased. Shifts in intensity, which may result from erythema or bolus-tissue air gaps, were detected most by %GP. However, neither metric detected beam-shape misalignment, such as that caused by dose to unintended areas, as well as currently employed metrics (Dice and MDC).

CONCLUSIONS: This study indicates that different radiotherapy incidents may be detected by comparing both inter- and intrafractional Cherenkov images with a corresponding image similarity metric, varying with the type of incident. Future work will involve determining appropriate thresholds per metric for automatic flagging.

ADVANCES IN KNOWLEDGE: Classifying different algorithms for the detection of various radiotherapy incidents allows for the development of an automatic flagging system, eliminating the burden of manual review of Cherenkov images.

来源:Br J Radiol,查看原文。版权归原作者所有。

引用

  1. Decker, Alexander, Bruza, et al. Performance comparison of quantitative metrics for analysis of in vivo Cherenkov imaging incident detection during radiotherapy. Br J Radiol. 2022. PMID: 35834415. doi:10.1259/bjr.20211346

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。