徒手三维超声系统测量肩部肌肉三维表面形状的效度验证
本研究以10名健康男性为对象,用定制固定支架保持体位,比较徒手三维超声与磁共振成像测量冈上肌、冈下肌和三角肌后束的体积与三维表面形状。结果显示两种模型的肌肉体积平均差异在-0.51 cm³以内,表面距离中位数小于1.12 mm,Dice相似系数大于0.85,提示在允许上述误差的前提下,三维超声可作为磁共振成像的替代方法测量肩部肌肉体积和表面形状。
要点
- 三维超声与磁共振成像所建肌肉模型的体积平均差异在-0.51 cm³以内。
- 三维表面形状的表面距离中位数小于1.12 mm,Dice相似系数大于0.85。
- 在允许上述误差的前提下,三维超声可作为磁共振成像的替代方法测量肩部肌肉体积和表面形状。
核心结论
- 在允许上述误差的前提下,三维超声可作为磁共振成像的替代方法测量肩部肌肉体积和表面形状。
研究设计与样本
- 效度验证研究,纳入10名健康男性,使用定制固定支架保持体位,以磁共振成像为参考标准,比较徒手三维超声与磁共振成像所建冈上肌、冈下肌和三角肌后束三维表面模型的体积和表面形状一致性。
诊断标准与关键数值
- 肌肉体积:两种模型平均差异在-0.51 cm³以内。
- 三维表面形状:两种模型最近点距离的中位数小于1.12 mm,Dice相似系数大于0.85。
局限性
- 原文指出三维超声的准确性主要在下肢肌肉以体积为指标进行过验证。
- 原文结论以「given the aforementioned error is permitted」为前提,即上述误差被允许时才可将三维超声作为磁共振成像的替代方法。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Freehand 3-D ultrasound (3DUS) system is a promising technique for accurately assessing muscle morphology. However, its accuracy has been validated mainly in terms of volume by examining lower limb muscles. This study was aimed at validating 3DUS in the measurements of 3-D surface shape and volume by comparing them with magnetic resonance imaging (MRI) measurements while ensuring the reproducibility of participant posture by focusing on the shoulder muscles. The supraspinatus, infraspinatus and posterior deltoid muscles of 10 healthy men were scanned using 3DUS and MRI while secured by an immobilization support customized for each participant. A 3-D surface model of each muscle was created from the 3DUS and MRI methods, and the agreement between them was assessed. For the muscle volume, the mean difference between the two models was within -0.51 cm3. For the 3-D surface shape, the distances between the closest points of the two models and the Dice similarity coefficient were calculated. The results indicated that the median surface distance was less than 1.12 mm and the Dice similarity coefficient was larger than 0.85. These results suggest that, given the aforementioned error is permitted, 3DUS can be used as an alternative to MRI in measuring volume and surface shape, even for the shoulder muscles.
引用
- Umehara, Fukuda, Konda, et al. Validity of Freehand 3-D Ultrasound System in Measurement of the 3-D Surface Shape of Shoulder Muscles. Ultrasound Med Biol. 2022. PMID: 35831210. doi:10.1016/j.ultrasmedbio.2022.06.001