乳腺超声容积扫查成像:扩大乳腺癌检出影像可及性的新视野
无超声经验的医学生接受不足2小时培训后,用容积扫查成像(VSI)协议和Butterfly iQ手持探头扫描170个可触及乳腺肿块,以当日标准超声为参考,VSI检出乳腺肿块的敏感度97%、特异度100%,与标准超声一致率97.6%(κ=0.95),所有超声表现为肿块的癌症检出率100%,BI-RADS评估一致率87%(κ=0.82)。
要点
- 无超声经验医学生经不足2小时培训即可完成乳腺VSI扫查,肿块检出敏感度97%、特异度100%。
- VSI与标准超声对乳腺肿块一致率97.6%(κ=0.95),BI-RADS评估一致率87%(κ=0.82)。
- 所有超声表现为肿块的癌症VSI检出率100%,提示可用于医疗资源不足地区扩大乳腺诊断影像可及性。
核心结论
- 乳腺VSI用于可触及肿块的评估,有望增加医疗资源不足地区诊断影像的可及性,可能减少乳腺癌诊断延迟,从而可能改善发病率与死亡率。
研究设计与样本
- 前瞻性对照研究,由无超声经验的医学生在不足2小时培训后用Butterfly iQ手持探头按VSI协议扫描170个可触及乳腺肿块,VSI视频由乳腺影像主治医师判读,并与当日标准超声检查对比。
关键影像表现/征象
- VSI检出乳腺肿块与标准超声一致率97.6%(Cohen κ=0.95,P<.0001)。
- VSI与标准超声对肿块特征高度一致,BI-RADS评估一致率87%(Cohen κ=0.82,P<.0001)。
诊断标准与关键数值
- VSI检出乳腺肿块的敏感度97%、特异度100%(以同日标准超声检查为参考)。
- 所有超声表现为肿块的癌症VSI检出率100%。
临床应用与阅片要点
- 可触及乳腺肿块可用无超声经验操作者经短时培训后按VSI协议扫查,由乳腺影像医师判读视频。
- VSI判读需对照标准超声,关注肿块特征与BI-RADS分类的一致性。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: The majority of people in the world lack basic access to breast diagnostic imaging resulting in delay to diagnosis of breast cancer. In this study, we tested a volume sweep imaging (VSI) ultrasound protocol for evaluation of palpable breast lumps that can be performed by operators after minimal training without prior ultrasound experience as a means to increase accessibility to breast ultrasound.
METHODS: Medical students without prior ultrasound experience were trained for less than 2 hours on the VSI breast ultrasound protocol. Patients presenting with palpable breast lumps for standard of care ultrasound examination were scanned by a trained medical student with the VSI protocol using a Butterfly iQ handheld ultrasound probe. Video clips of the VSI scan imaging were later interpreted by an attending breast imager. Results of VSI scan interpretation were compared to the same-day standard of care ultrasound examination.
RESULTS: Medical students scanned 170 palpable lumps with the VSI protocol. There was 97% sensitivity and 100% specificity for a breast mass on VSI corresponding to 97.6% agreement with standard of care (Cohen's κ = 0.95, P < .0001). There was a detection rate of 100% for all cancer presenting as a sonographic mass. High agreement for mass characteristics between VSI and standard of care was observed, including 87% agreement on Breast Imaging-Reporting and Data System assessments (Cohen's κ = 0.82, P < .0001).
CONCLUSIONS: Breast ultrasound VSI for palpable lumps offers a promising means to increase access to diagnostic imaging in underserved areas. This approach could decrease delay to diagnosis for breast cancer, potentially improving morbidity and mortality.
引用
- Marini, Castaneda, Iyer, et al. Breast Ultrasound Volume Sweep Imaging: A New Horizon in Expanding Imaging Access for Breast Cancer Detection. J Ultrasound Med. 2022. PMID: 35802491. doi:10.1002/jum.16047
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。