三维超快超声定位显微成像用于冠状动脉血流评估
在离体跳动大鼠心脏(N=6)中,采用超声定位显微成像实现全心脏三维冠状动脉微血管成像(分辨率<20 μm,帧率>1000帧/秒),评估正常、腺苷输注及冠状动脉闭塞条件下的微血管解剖、流速与流量,并与微计算机断层扫描对比;在体经胸成像在N=3只大鼠中验证可行性。
要点
- 三维冠状动脉超声定位显微成像可在跳动心脏中显示分辨率<20 μm的微血管结构,流速从微小动脉10 mm/s至大动脉超过300 mm/s。
- 流量-半径关系符合幂律,指数为2.61(r2=0.96;P<0.001),与血管树标度律的理论预测和实验验证一致。
- 腺苷使微血管冠状动脉流量增加2倍,与主动脉总灌注流量从8.80±1.03 mL/min增至16.54±2.35 mL/min(P<0.001)吻合。
核心结论
- 三维冠状动脉超声定位显微成像为跳动心脏微血管水平的冠状动脉解剖与功能提供了前所未有的信息,该新技术具有高度转化潜力,有潜力成为冠状动脉微循环临床研究的主要工具。
研究设计与样本
- 研究在离体跳动大鼠心脏(N=6)中行三维冠状动脉超声定位显微成像,使用超声造影剂(Sonovue,Bracco)和超声矩阵换能器连接高通道数超快电子系统;在正常条件、腺苷输注血管扩张及冠状动脉闭塞下评估三维冠状动脉微血管解剖、流速和流量;冠状动脉血管树与固定心脏的微计算机断层扫描对比;在麻醉大鼠(N=3)中评估在体经胸冠状动脉超声定位显微成像。
关键影像表现/征象
- 三维冠状动脉超声定位显微成像可在跳动心脏中三维显示冠状动脉血管树,尺度低至微血管结构(分辨率<20 μm)。
- 绝对流速估计范围从微小动脉的10 mm/s至大动脉的超过300 mm/s。
- 流量-半径关系拟合幂律的指数为2.61(r2=0.96;P<0.001),与血管树标度律的理论预测和实验验证一致。
- 腺苷作用下微血管冠状动脉流量增加2倍,与主动脉总灌注流量从8.80±1.03 mL/min增至16.54±2.35 mL/min(P<0.001)吻合。
局限性
- 在体经胸冠状动脉超声定位显微成像的可行性仅在N=3只大鼠中验证。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Direct assessment of the coronary microcirculation has long been hampered by the limited spatial and temporal resolutions of cardiac imaging modalities.
OBJECTIVES: The purpose of this study was to demonstrate 3-dimensional (3D) coronary ultrasound localization microscopy (CorULM) of the whole heart beyond the acoustic diffraction limit (<20 μm resolution) at ultrafast frame rate (>1000 images/s).
METHODS: CorULM was performed in isolated beating rat hearts (N = 6) with ultrasound contrast agents (Sonovue, Bracco), using an ultrasonic matrix transducer connected to a high channel-count ultrafast electronics. We assessed the 3D coronary microvascular anatomy, flow velocity, and flow rate of beating hearts under normal conditions, during vasodilator adenosine infusion, and during coronary occlusion. The coronary vasculature was compared with micro-computed tomography performed on the fixed heart. In vivo transthoracic CorULM was eventually assessed on anaesthetized rats (N = 3).
RESULTS: CorULM enables the 3D visualization of the coronary vasculature in beating hearts at a scale down to microvascular structures (<20 μm resolution). Absolute flow velocity estimates range from 10 mm/s in tiny arterioles up to more than 300 mm/s in large arteries. Fitting to a power law, the flow rate-radius relationship provides an exponent of 2.61 (r2 = 0.96; P < 0.001), which is consistent with theoretical predictions and experimental validations of scaling laws in vascular trees. A 2-fold increase of the microvascular coronary flow rate is found in response to adenosine, which is in good agreement with the overall perfusion flow rate measured in the aorta (control measurement) that increased from 8.80 ± 1.03 mL/min to 16.54 ± 2.35 mL/min (P < 0.001). The feasibility of CorULM was demonstrated in vivo for N = 3 rats.
CONCLUSIONS: CorULM provides unprecedented insights into the anatomy and function of coronary arteries at the microvasculature level in beating hearts. This new technology is highly translational and has the potential to become a major tool for the clinical investigation of the coronary microcirculation.
引用
- Demeulenaere, Sandoval, Mateo, et al. Coronary Flow Assessment Using 3-Dimensional Ultrafast Ultrasound Localization Microscopy. JACC Cardiovasc Imaging. 2022. PMID: 35798395. doi:10.1016/j.jcmg.2022.02.008