基于256层CT将脑CT灌注与头颈部CT血管成像联合的新「一站式」扫描方案用于卒中患者的研究
2020年3月至8月纳入60例疑似急性卒中患者,随机分为两组,A组采用传统分离的脑CTP与头颈部CTA方案,B组采用将头颈部CTA数据插入脑CTP动脉期峰值时间的新「一站式」方案。B组对比剂剂量60 mL较A组100 mL降低40%,成像时间52.5±2.6 s较A组74.9±3.3 s降低约43%,图像质量定量与定性均无显著差异(均P>0.05),B组剂量长度乘积轻度降低(1139.0±45.3 vs 1211.6±31.9 mGy·cm,P<0.001)。
要点
- 新「一站式」方案将头颈部CTA数据插入脑CTP动脉期峰值时间,插入点由试验性团注确定。
- B组对比剂60 mL较A组100 mL降低40%,成像时间缩短约43%(52.5±2.6 s vs 74.9±3.3 s)。
- 两组图像质量定量与定性均无显著差异(均P>0.05),B组剂量长度乘积轻度降低(P<0.001)。
核心结论
- 在256层CT上,所提出的联合脑CTP与头颈部CTA的「一站式」扫描方案较传统分离检查方案可缩短成像时间、降低对比剂剂量,且不影响图像质量或灌注结果。
研究设计与样本
- 随机对照研究,纳入2020年3月至8月60例(男30例、女30例)年龄22-88岁疑似急性卒中患者,随机分为2组,使用256层CT系统行脑CTP与头颈部CTA检查。
诊断标准与关键数值
- A组采用传统方案,分别行脑CTP与头颈部CT检查(含平扫与增强采集);B组采用新「一站式」方案,将头颈部CTA数据插入脑CTP动脉期峰值时间(PT),插入点由试验性团注确定。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: We sought to evaluate the performance of a new "one-stop-shop" scan protocol combining brain computed tomography perfusion (CTP) and head-and-neck CT angiography (CTA) imaging for acute stroke patients using a 256-detector CT scanner.
METHOD: From March to August 2020, 60 patients (30 men and 30 women) aged 22-88 years with suspected acute stroke were enrolled and randomly divided into 2 groups to undergo brain CTP and head-and-neck CTA with a 256-detector CT system. Group A used traditional scan protocol with a separate brain CTP and head-and-neck CT examination that included non-contrast-enhanced and contrast-enhanced acquisitions; group B used the new "one-stop-shop" scan protocol with head-and-neck CTA data inserted into brain CTP scans at the peak time (PT) of the arterial phase. The insertion point of the head-and-neck CTA data was determined by a test bolus. The examination time, contrast dose, radiation dose, and image quality were compared between the groups.
RESULTS: The total contrast dose was reduced by 40% in group B compared to group A (60 mL vs. 100 mL). The imaging time was 52.5 ± 2.6 s in group B and 74.9 ± 3.3 s in group A, showing a reduction of approximately 43% in group B. There was no significant difference in image quality both quantitatively and qualitatively between the groups (all P > 0.05). Group B had a slight reduction in dose length product (1139.0 ± 45.3 vs. 1211.6 ± 31.9 mGy·cm, P < 0.001).
CONCLUSIONS: The proposed "one-stop-shop" scan protocol combining brain CTP and head-and-neck CTA on a 256-detector CT system can reduce imaging time and contrast dose, without affecting image quality or perfusion results, compared to the traditional protocol of separating the examinations.
引用
- Mo, Cui, Yuan, et al. Study on a new "One-stop-shop" scan protocol combining brain CT perfusion and head-and-neck CT angiography by using 256-detector CT for stroke patients. Eur J Radiol. 2022. PMID: 35797790. doi:10.1016/j.ejrad.2022.110426