CT引导下颅底与颅颈交界区穿刺的机架-针-靶对齐技术
回顾性分析39例CT引导下颅底及寰枢椎穿刺操作,均采用机架倾斜联合机架-针-靶对齐技术,其中9例为非传统入路。2例发生围手术期并发症,1例无症状、1例24小时内缓解,所有操作均成功完成并准确到达靶点。
要点
- 机架倾斜联合机架-针-靶对齐技术可获得双斜位穿刺轨迹,用于颅底和寰枢椎穿刺。
- 2例围手术期并发症:1例无症状,1例24小时内缓解;所有操作成功到达靶点。
核心结论
- 机架倾斜与机架-针-靶对齐技术使CT引导下颅底和寰枢椎穿刺可获得双斜位轨迹,减少穿刺轨迹不确定性,在传统和非传统入路中均实现安全置针。
研究设计与样本
- 回顾性分析连续39例CT引导下颅底及寰枢椎穿刺操作,包括细针抽吸和粗针活检组织取样、寰枢椎肿瘤性溶骨病变骨水泥成形术、疼痛治疗注射及硬膜穿刺脑脊液取样,均采用机架-针-靶对齐技术,记录操作并发症。
临床应用与阅片要点
- 对颅底和寰枢椎病变行CT引导穿刺时,可利用机架倾斜联合机架-针-靶对齐技术获得双斜位轨迹,以匹配理想的头尾倾斜穿刺路径。
- 传统入路包括经上颌骨、后外侧和颧下入路,非传统入路包括乳突下、颧上、经鼻、经乳突和经耳入路。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: CT-guided percutaneous procedures involving the skull base and atlanto-axial cervical spine pose particular challenges due to high density of vital vascular and nervous structures and because the ideal needle trajectory often has a cranio-caudal obliquity different from the axial scan plane. We describe how the variable CT gantry tilt, combined with gantry-needle-target alignment technique, is used to obtain precise and safe needle placement in conventional and non-conventional approaches to the skull base and the atlanto-axial spine.
METHODS: We retrospectively analyzed consecutive CT-guided needle accesses to the skull base and atlanto-axial spine performed for tissue sampling through fine-needle aspirates and core biopsies, cementoplasty of neoplastic lytic lesions of atlanto-axial spine, pain management injections, and dural puncture for cerebro-spinal fluid sampling. All the accesses were performed with the gantry-needle-target alignment technique. Procedural complications were recorded.
RESULTS: Thirty-nine CT-guided procedures were analyzed. Paramaxillary approach was used in 15 cases, postero-lateral in 11, subzygomatic in 3. Nine non-conventional approach were performed: submastoid in 3 cases, suprazygomatic in 2, trans-nasal in 2, trans-mastoid in 1, and trans-auricular in 1. Two peri-procedural complications occurred: one asymptomatic and one resolved within 24 h. All the procedures were successfully completed with successful needle access to the target.
CONCLUSION: The gantry tilt and gantry-needle-target alignment technique allows to obtain double-oblique needle accesses for CT-guided procedures involving the skull base and atlanto-axial cervical spine, minimizing uncertainty of needle trajectory and obtaining safe needle placement in conventional and non-conventional approaches.
引用
- Pileggi, Ventura, Di Napoli, et al. Gantry-needle-target alignment technique for CT-guided needle approaches to the skull base and cranio-cervical junction. Neuroradiology. 2022. PMID: 35788697. doi:10.1007/s00234-022-03005-9
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。