超声肝脏脂肪定量测量的观察者间变异性

Interobserver Variability in Ultrasound-Based Liver Fat Quantification.
PubMed 文献 · J Ultrasound Med · 2022-07-02

这项前瞻性研究纳入100例已知或疑似非酒精性脂肪性肝病的患者,由两名互盲观察者独立进行组织衰减成像(TAI)和组织散射分布成像(TSI)测量,每种技术各测5次取中位数。结果显示TAI的ICC为0.970、TSI的ICC为0.938,观察者间一致性均极好;两种技术所需额外时间较少。

要点

  • TAI观察者间一致性ICC为0.970,TSI为0.938,均属极好。
  • 观察者1与观察者2的TAI中位数分别为0.75和0.74 dB/cm/MHz。
  • TAI平均耗时55.1±7.8秒与59.9±6.6秒,TSI为49.1±5.8秒与54.1±5.4秒。

核心结论

  • TAI和TSI两种技术均具有高度可重复性,可在日常实践中实施且所需额外时间较少。

研究设计与样本

  • 前瞻性研究,纳入100例已知或疑似非酒精性脂肪性肝病患者,两名观察者互盲独立操作,使用组织衰减成像(TAI)和组织散射分布成像(TSI)两种定量超声技术,每种技术各获取5次测量并记录中位数。

关键影像表现/征象

  • TAI测量中位数:观察者1为0.75 dB/cm/MHz,观察者2为0.74 dB/cm/MHz。
  • TSI测量中位数:观察者1为93.53,观察者2为92.58。
  • 两名观察者均对肝脂肪变性进行视觉评估,并采用Spearman相关检验分析定量超声测量值与视觉肝脂肪变性分级之间的相关性。

诊断标准与关键数值

  • TAI观察者间一致性ICC为0.970,判为极好。
  • TSI观察者间一致性ICC为0.938,判为极好。

临床应用与阅片要点

  • TAI平均耗时:观察者1为55.1±7.8秒,观察者2为59.9±6.6秒。
  • TSI平均耗时:观察者1为49.1±5.8秒,观察者2为54.1±5.4秒。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To assess interobserver variability in ultrasound-based quantitative liver fat content measurements and to determine how much time these quantitative ultrasound (QUS) techniques require.

METHODS: One hundred patients with known or suspected of having nonalcoholic fatty liver disease were included in this prospective study. Two observers who were blinded to each other measurements performed tissue attenuation imaging (TAI) and tissue scatter distribution imaging (TSI) techniques independently. Both observers assessed hepatic steatosis visually and obtained 5 measurements for each QUS technique and the median values of the measurements were recorded. Spearman's correlation test was used to assess the correlation between QUS measurements and visual hepatic stetaosis grades. Intraclass correlation coefficient (ICC) test was used to assess interobserver variability in QUS measurements.

RESULTS: The median values of TAI measurements for the observers 1 and 2 were 0.75 and 0.74 dB/cm/MHz, respectively. The median values of TSI measurements for the observers 1 and 2 were 93.53 and 92.58, respectively. The interobserver agreement in TAI (ICC: 0.970) and TSI (ICC: 0.938) measurements were excellent. The mean of the required time period for TAI technique were 55.1 ± 7.8 and 59.9 ± 6.6 seconds for the observers 1 and 2, respectively. The mean of the required time period for TSI technique were 49.1 ± 5.8 and 54.1 ± 5.4 seconds for the observers 1 and 2, respectively.

CONCLUSION: The current study revealed that both TAI and TSI techniques are highly reproducible and can be implemented into daily practice with little additional time requirement.

来源:J Ultrasound Med,查看原文。版权归原作者所有。

引用

  1. Şendur, Cerit, Ibrahimkhanli, et al. Interobserver Variability in Ultrasound-Based Liver Fat Quantification. J Ultrasound Med. 2022. PMID: 35778902. doi:10.1002/jum.16048

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。