文献速递头颈神经CT

颞骨超高分辨率光子计数与能量积分探测器CT中锡预滤过光谱塑形的比较研究

Spectral Shaping Via Tin Prefiltration in Ultra-High-Resolution Photon-Counting and Energy-Integrating Detector CT of the Temporal Bone.
PubMed 文献 · Invest Radiol · 2022-06-24

研究在5具尸头标本上比较了光子计数探测器CT(PCD-CT)与能量积分探测器CT(EID-CT)行颞骨超高分辨率成像时锡预滤过光谱塑形的剂量节省潜力,PCD-CT图像质量优于EID-CT并可实现平均79.3%±3.9%的剂量降低,锡预滤过在EID-CT中可额外节省剂量而在PCD-CT中因固有降噪优势剂量节省潜力降低。

要点

  • PCD-CT在低剂量(171.2±10.3 HU)和中剂量(134.7±4.5 HU)下图像噪声低于EID-CT全剂量(177.0±14.2 HU)。
  • 在120 kVp下PCD-CT达到与EID-CT匹配图像噪声时,平均剂量降低79.3%±3.9%。
  • PCD-CT各剂量水平的主观图像质量均优于EID-CT,且低剂量EID-CT多数被判为不可诊断而同等剂量PCD-CT获满意以上评级。

核心结论

  • 颞骨超高分辨率检查中PCD-CT图像质量优于EID-CT并可显著节省剂量;锡预滤过高电压方案可在EID-CT中额外节省剂量,而PCD-CT因固有降噪优势使光谱塑形的剂量降低潜力减小。

研究设计与样本

  • 在5具尸头标本上,使用PCD-CT和EID-CT系统以匹配剂量方案(低/中/全剂量:4.8/7.6-7.7/27.0-27.1 mGy)并分别配合或不配合锡预滤过,共完成12次超高分辨率检查。

关键影像表现/征象

  • 标准成像采用120 kVp,光谱塑形方案采用锡预滤过可用的最高管电压(EID-CT:Sn 150 kVp;PCD-CT:Sn 140 kVp)。
  • 不论是否使用锡预滤过,PCD-CT低剂量(171.2±10.3 HU)和中剂量(134.7±4.5 HU)图像噪声均低于EID-CT全剂量(177.0±14.2 HU;P<0.001)。
  • Sn 150 kVp EID-CT在高剂量和中剂量水平图像质量优于120 kVp EID-CT,但差异未达统计学显著(P>0.050)。
  • 剂量匹配的PCD-CT有无锡预滤过之间图像质量无差异(P≥0.928)。

诊断标准与关键数值

  • 以120 kVp EID-CT匹配图像噪声为目标,120 kVp PCD-CT可实现平均剂量降低79.3%±3.9%。
  • 三位放射科医师独立评估图像质量,评估者间一致性组内相关系数为0.903(优秀)。
  • PCD-CT各剂量水平主观图像质量均优于EID-CT(P<0.050)。

临床应用与阅片要点

  • 颞骨超高分辨率成像中PCD-CT可在更低剂量下获得优于EID-CT的图像质量。
  • EID-CT采用锡预滤过高电压方案有助于额外节省剂量。
  • PCD-CT因固有降噪优势,锡预滤过光谱塑形带来的额外剂量节省潜力减小。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: Hardening the x-ray beam, tin prefiltration is established for imaging of high-contrast subjects in energy-integrating detector computed tomography (EID-CT). With this work, we aimed to investigate the dose-saving potential of spectral shaping via tin prefiltration in photon-counting detector CT (PCD-CT) of the temporal bone.

METHODS: Deploying dose-matched scan protocols with and without tin prefiltration on a PCD-CT and EID-CT system (low-/intermediate-/full-dose: 4.8/7.6-7.7/27.0-27.1 mGy), 12 ultra-high-resolution examinations were performed on each of 5 cadaveric heads. While 120 kVp was applied for standard imaging, the protocols with spectral shaping used the highest potential available with tin prefiltration (EID-CT: Sn 150 kVp, PCD-CT: Sn 140 kVp). Contrast-to-noise ratios and dose-saving potential by spectral shaping were computed for each scanner. Three radiologists independently assessed the image quality of each examination with the intraclass correlation coefficient being computed to measure interrater agreement.

RESULTS: Regardless of tin prefiltration, PCD-CT with low (171.2 ± 10.3 HU) and intermediate radiation dose (134.7 ± 4.5 HU) provided less image noise than full-dose EID-CT (177.0 ± 14.2 HU; P < 0.001). Targeting matched image noise to 120 kVp EID-CT, mean dose reduction of 79.3% ± 3.9% could be realized in 120 kVp PCD-CT. Subjective image quality of PCD-CT was better than of EID-CT on each dose level ( P < 0.050). While no distinction was found between dose-matched PCD-CT with and without tin prefiltration ( P ≥ 0.928), Sn 150 kVp EID-CT provided better image quality than 120 kVp EID-CT at high and intermediate dose levels ( P > 0.050). The majority of low-dose EID-CT examinations were considered not diagnostic, whereas PCD-CT scans of the same dose level received satisfactory or better ratings. Interrater reliability was excellent (intraclass correlation coefficient 0.903).

CONCLUSIONS: PCD-CT provides superior image quality and significant dose savings compared with EID-CT for ultra-high-resolution examinations of the temporal bone. Aiming for matched image noise, high-voltage scan protocols with tin prefiltration facilitate additional dose saving in EID-CT, whereas superior inherent denoising decreases the dose reduction potential of spectral shaping in PCD-CT.

来源:Invest Radiol,查看原文。版权归原作者所有。

引用

  1. Grunz, Heidenreich, Lennartz, et al. Spectral Shaping Via Tin Prefiltration in Ultra-High-Resolution Photon-Counting and Energy-Integrating Detector CT of the Temporal Bone. Invest Radiol. 2022. PMID: 35776435. doi:10.1097/RLI.0000000000000901