AI-CAD用于鉴别乳腺X线摄影上仅表现为钙化的病变:结合ACR BI-RADS钙化描述词的结果分析
研究纳入2016年6月至2018年5月间500例女性535个仅表现为钙化的乳腺X线病灶,评估AI-CAD对钙化的分诊能力并与经验丰富的乳腺放射科医生比较。结果显示,相同形态或BI-RADS评估的钙化中,AI-CAD评分阳性者PPV显著更高;AI-CAD与放射科医生诊断性能无显著差异。
要点
- 相同形态或BI-RADS评估的钙化中,AI-CAD评分阳性者PPV显著更高。
- AI-CAD与经验丰富放射科医生对仅钙化病灶的诊断性能无显著差异。
- 恶性钙化中,AI-CAD评分阳性者浸润癌比例高于阴性者,但差异无统计学意义(p>0.05)。
核心结论
- 在形态或BI-RADS评估相同的钙化中,AI-CAD评分阳性者PPV显著更高;AI-CAD对仅钙化病灶的诊断性能与经验丰富的放射科医生相似。
研究设计与样本
- 回顾性研究纳入2016年6月至2018年5月常规乳腺X线摄影中500例女性(平均年龄48.8岁)的535个仅表现为钙化的病灶,并加摄放大位;由一名专职乳腺放射科医生按ACR BI-RADS评估形态、分布和最终评估分类;AI-CAD分析常规乳腺X线摄影,提供AI-CAD标记和0%至100%的AI-CAD评分,评分≥10%视为阳性;恶性或良性金标准经组织病理学诊断或至少1年影像随访确认。
诊断标准与关键数值
- AI-CAD评分≥10%为阳性;535个钙化中215个(40.2%)为恶性;AI-CAD与经验丰富放射科医生比较,敏感度92.1% vs 95.4%(p=0.125)、特异度71.9% vs 72.5%(p=0.842)、准确率80.0% vs 81.7%(p=0.413),均无显著差异。
临床应用与阅片要点
- 对于形态或BI-RADS评估相同的钙化,可结合AI-CAD评分阳性(≥10%)来识别PPV更高的病灶。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To evaluate how AI-CAD triages calcifications and to compare its performance to an experienced breast radiologist.
METHODS: Among routine mammography performed between June 2016 and May 2018, 535 lesions detected as calcifications only on mammography in 500 women (mean age, 48.8 years) that were additionally interpreted with additional magnification views were included in this study. One dedicated breast radiologist retrospectively reviewed the magnification mammograms to assess morphology, distribution, and final assessment category according to ACR BI-RADS. AI-CAD analyzed routine mammograms providing AI-CAD marks and corresponding AI-CAD scores (ranging from 0 to 100%), for which values ≥ 10% were considered positive. Ground truth in terms of malignancy or benignity was confirmed with a histopathologic diagnosis or at least 1 year of imaging follow - up.
RESULTS: Of the 535 calcifications, 215 (40.2%) were malignant. Calcifications with positive AI-CAD scores showed significantly higher PPVs compared to calcifications with negative scores for all morphology (all p < 0.05). PPVs were significantly higher in calcifications with positive AI-CAD scores compared to those with negative scores for BI-RADS 3, 4a, or 4b assessments (all p < 0.05). AI-CAD and the experienced radiologist did not show significant difference in diagnostic performance; sensitivity 92.1% vs 95.4% (p = 0.125), specificity 71.9% vs 72.5% (p = 0.842), and accuracy 80.0% vs 81.7% (p = 0.413).
CONCLUSION: Among calcifications with same morphology or BI-RADS assessment, those with positive AI-CAD scores had significantly higher PPVs. AI-CAD showed similar diagnostic performances to the experienced radiologist for calcifications detected on mammography.
KEY POINTS: • Among calcifications with same morphology or BI-RADS assessment, those with positive AI-CAD scores had significantly higher PPVs. • AI-CAD showed similar diagnostic performance to an experienced radiologist in assessing lesions detected as calcifications only on mammography. • Among malignant calcifications, calcifications with positive AI-CAD scores showed higher rates of invasive cancers than calcifications with negative scores (all p > 0.05).
引用
- Yoon, Lee, Kim, et al. AI-CAD for differentiating lesions presenting as calcifications only on mammography: outcome analysis incorporating the ACR BI-RADS descriptors for calcifications. Eur Radiol. 2022. PMID: 35748900. doi:10.1007/s00330-022-08961-7
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。