文献速递神经外周血管MR

7T MRI期间钛合金脑动脉瘤夹加热评估

Assessment of Heating on Titanium Alloy Cerebral Aneurysm Clips during 7T MRI.
PubMed 文献 · AJNR Am J Neuroradiol · 2022-06-23

本研究在体模中测试5种钛合金动脉瘤夹在7T MRI下的升温情况,结果显示直角夹和短直夹各部位温升均低于1°C,而45°角夹、直窗孔夹和长直夹部分部位温升超过1°C,长直夹所有部位温升超过2°C。

要点

  • 直角夹和短直夹在7T MRI下各测点温升均低于1°C。
  • 45°角夹 angled 部分和直窗孔夹头端温升超过1°C。
  • 长直夹所有测点温升均超过2°C。

核心结论

  • 直角夹和短直夹在7T MRI期间温升低于监管限值,而45°角夹、直窗孔夹和长直夹温升超过该限值。

研究设计与样本

  • 体模实验研究,测试5种钛合金动脉瘤夹(短、长、直、成角、窗孔型的组合),置于填充凝胶生理盐水与聚丙烯酸混合物的体模中,行7T MRI 3D T1WI稳态 spoiled gradient recalled 采集,用MR兼容光纤温度计测温。

关键影像表现/征象

  • 直角夹和短直夹所有测点温升均低于1°C。
  • 45°角夹的成角部位温升超过1°C。
  • 直窗孔夹的头端温升超过1°C。
  • 长直夹所有测点温升均超过2°C。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND AND PURPOSE: Patients with cerebral aneurysms often undergo MR imaging after microsurgical clipping. Ultra-high-field MR imaging at 7T may provide high diagnostic capability in such clinical situations. However, titanium alloy clips have safety issues such as adverse interactions with static magnetic fields and radiofrequency-induced heating during 7T MR imaging. The purpose of this study was to quantitatively assess temperature increases on various types of titanium alloy aneurysm clips during 7T MR imaging.

MATERIALS AND METHODS: Five types of titanium alloy aneurysm clips were tested, including combinations of short, long, straight, angled, and fenestrated types. Each clip was set in a phantom filled with gelled saline mixed with polyacrylic acid and underwent 7T MR imaging with 3D T1WI with a spoiled gradient recalled acquisition in the steady-state technique. Temperature was chronologically measured at the tips of the clip blade and head, angled part of the clip, and 5 mm from the tip of the clip head using MR imaging-compatible fiber-optic thermometers.

RESULTS: Temperature increases at all locations for right-angled and short straight clips were <1°C. Temperature increases at the angled part for the 45° angled clip and the tip of the clip head for the straight fenestrated clip were >1°C. Temperature increases at all locations for the long straight clip were >2°C.

CONCLUSIONS: Temperature increases on the right-angled and short straight clips remained below the regulatory limit during 7T MR imaging, but temperature increases on the 45° angled, straight fenestrated, and long straight clips exceeded this limit.

来源:AJNR Am J Neuroradiol,查看原文。版权归原作者所有。