文献速递胸部技术与安全CT

肺优化深度学习重建在超低剂量CT中的应用

Lung-Optimized Deep-Learning-Based Reconstruction for Ultralow-Dose CT.
PubMed 文献 · Acad Radiol · 2022-06-20

该研究在体模和14例非肥胖患者中评估肺专用深度学习重建(DLR)在超低剂量CT(ULDCT)中的图像特性,并与混合迭代重建(HIR)和基于模型的迭代重建(MBIR)比较。结果显示ULDCT-DLR的图像噪声低于、CNR高于LDCT-1、LDCT-2和ULDCT-HIR(p<0.01),整体质量高于ULDCT-HIR和ULDCT-MBIR(p<0.01),与LDCT-2几乎相当(平均评分3.4±0.5);临床ULDCT中DLR的主观可接受性高于HIR和MBIR(p<0.01)。结论认为肺优化DLR改善图像质量,并提供可能比HIR和MBIR更大的剂量优化机会。

要点

  • ULDCT-DLR图像噪声低于LDCT-1、LDCT-2和ULDCT-HIR,CNR高于上述三者(p<0.01)。
  • ULDCT-DLR整体质量高于ULDCT-HIR和ULDCT-MBIR(p<0.01),与LDCT-2几乎相当(平均评分3.4±0.5)。
  • 临床ULDCT中DLR主观可接受性高于HIR和MBIR(p<0.01)。

核心结论

  • 肺优化DLR改善图像质量,并提供可能比HIR和MBIR更大的剂量优化机会。

研究设计与样本

  • 体模研究加临床图像评估:在320排扫描仪上以50 mA(LDCT-1)、25 mA(LDCT-2)和10 mA(ULDCT)扫描 anthropomorphic 胸部体模;LDCT用HIR重建,ULDCT分别用HIR(ULDCT-HIR)、MBIR(ULDCT-MBIR)和DLR(ULDCT-DLR)重建;以LDCT图像为参考标准对ULDCT进行5分制主观评分;另用物理评估体模在7个剂量下扫描以评估NPS和TTF;临床部分纳入14例非肥胖患者的10 mA ULDCT图像,用HIR、MBIR和DLR重建并排序主观可接受性。

诊断标准与关键数值

  • 主观评分采用5分制:1=明显劣于LDCT-2,3=与LDCT-2相当,5=与LDCT-1相当。
  • ULDCT-DLR整体质量平均评分3.4±0.5,与LDCT-2几乎相当。
  • DLR在高对比度物体上获得最高NPS峰值频率和TTF50%。

局限性

  • 临床部分仅纳入14例非肥胖患者。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

RATIONALE AND OBJECTIVES: To evaluate the image properties of lung-specialized deep-learning-based reconstruction (DLR) and its applicability in ultralow-dose CT (ULDCT) relative to hybrid- (HIR) and model-based iterative-reconstructions (MBIR).

MATERIALS AND METHODS: An anthropomorphic chest phantom was scanned on a 320-row scanner at 50-mA (low-dose-CT 1 [LDCT-1]), 25-mA (LDCT-2), and 10-mA (ULDCT). LDCT were reconstructed with HIR; ULDCT images were reconstructed with HIR (ULDCT-HIR), MBIR (ULDCT-MBIR), and DLR (ULDCT-DLR). Image noise and contrast-to-noise ratio (CNR) were quantified. With the LDCT images as reference standards, ULDCT image qualities were subjectively scored on a 5-point scale (1 = substantially inferior to LDCT-2, 3 = comparable to LDCT-2, 5 = comparable to LDCT-1). For task-based image quality analyses, a physical evaluation phantom was scanned at seven doses to achieve the noise levels equivalent to chest phantom; noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated. Clinical ULDCT (10-mA) images obtained in 14 nonobese patients were reconstructed with HIR, MBIR, and DLR; the subjective acceptability was ranked.

RESULTS: Image noise was lower and CNR was higher in ULDCT-DLR and ULDCT-MBIR than in LDCT-1, LDCT-2, and ULDCT-HIR (p < 0.01). The overall quality of ULDCT-DLR was higher than of ULDCT-HIR and ULDCT-MBIR (p < 0.01), and almost comparable with that of LDCT-2 (mean score: 3.4 ± 0.5). DLR yielded the highest NPS peak frequency and TTF50% for high-contrast object. In clinical ULDCT images, the subjective acceptability of DLR was higher than of HIR and MBIR (p < 0.01).

CONCLUSION: DLR optimized for lung CT improves image quality and provides possible greater dose optimization opportunity than HIR and MBIR.

来源:Acad Radiol,查看原文。版权归原作者所有。

引用

  1. Goto, Nagayama, Sakabe, et al. Lung-Optimized Deep-Learning-Based Reconstruction for Ultralow-Dose CT. Acad Radiol. 2022. PMID: 35738988. doi:10.1016/j.acra.2022.04.025

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。