4D flow MRI评估肝动脉血流动力学:运动与空间分辨率相关误差的体外分析及20名志愿者体内可行性研究
本研究通过肝动脉动态体模和20名连续志愿者,在3 T下评估笛卡尔4D flow序列不同空间分辨率(0.5–1 mm各向同性)和导航门控接受窗(±8至±2 mm)及2D相位对比序列的准确性。体外显示提高空间分辨率比缩窄导航窗更能降低误差;体内高分辨率序列血管显示率更高(90% vs 71%),观察者间一致性相似但变异性更低,4D flow在肝动脉分叉处流入流出差异小于2D相位对比。结论认为高分辨率4D flow可评估肝动脉解剖和血流动力学,准确性、血管可见性、观察者间信度和内部一致性均改善。
要点
- 高分辨率4D flow序列显示肝动脉和胃十二指肠动脉的比例为90%,高于低分辨率序列的71%。
- 高分辨率序列的观察者间一致性κ=0.660,低分辨率序列κ=0.704,两者相似。
- 4D flow在肝动脉分叉处流入流出差异为11.03±5.05%和15.69±6.14%,低于2D相位对比的28.77±21.01%。
核心结论
- 高分辨率4D flow可评估肝动脉解剖和血流动力学,准确性、血管可见性、观察者间信度和内部一致性均改善。
研究设计与样本
- 体外研究采用肝动脉动态体模,体内研究纳入20名连续志愿者,在3 T下扫描。
- 体外评估笛卡尔4D flow序列(k空间重排序和导航门控)在四种空间分辨率(0.5–1 mm各向同性)和导航接受窗(±8至±2 mm)下的准确性,并与一个2D相位对比序列(平面内0.5 mm)比较。
- 体内评估两个以胃十二指肠动脉和肝动脉分支为中心的序列的观察者内和观察者间一致性,并与2D相位对比比较。
诊断标准与关键数值
- 体外:较高空间分辨率比较窄导航窗导致更大的误差降低(流量误差从30.5%降至-4.67% vs 从-6.64%降至-4.67%)。
- 体内:高分辨率序列显示肝动脉和胃十二指肠动脉的比例为90%,低分辨率序列为71%。
- 观察者间一致性:高分辨率序列κ=0.660,低分辨率序列κ=0.704。
- 高分辨率序列的变异系数:面积CV=20.04% vs 30.67%,流量CV=34.92% vs 51.99%,平均速度CV=26.47% vs 44.76%。
- 4D flow在肝动脉分叉处流入流出差异为11.03±5.05%和15.69±6.14%,2D相位对比为28.77±21.01%。
临床应用与阅片要点
- 原文未据此给出推荐。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To assess the ability of four-dimensional (4D) flow MRI to measure hepatic arterial hemodynamics by determining the effects of spatial resolution and respiratory motion suppression in vitro and its applicability in vivo with comparison to two-dimensional (2D) phase-contrast MRI.
METHODS: A dynamic hepatic artery phantom and 20 consecutive volunteers were scanned. The accuracies of Cartesian 4D flow sequences with k-space reordering and navigator gating at four spatial resolutions (0.5- to 1-mm isotropic) and navigator acceptance windows (± 8 to ± 2 mm) and one 2D phase-contrast sequence (0.5-mm in -plane) were assessed in vitro at 3 T. Two sequences centered on gastroduodenal and hepatic artery branches were assessed in vivo for intra - and interobserver agreement and compared to 2D phase-contrast.
RESULTS: In vitro, higher spatial resolution led to a greater decrease in error than narrower navigator window (30.5 to -4.67% vs -6.64 to -4.67% for flow). In vivo, hepatic and gastroduodenal arteries were more often visualized with the higher resolution sequence (90 vs 71%). Despite similar interobserver agreement (κ = 0.660 and 0.704), the higher resolution sequence had lower variability for area (CV = 20.04 vs 30.67%), flow (CV = 34.92 vs 51.99%), and average velocity (CV = 26.47 vs 44.76%). 4D flow had lower differences between inflow and outflow at the hepatic artery bifurcation (11.03 ± 5.05% and 15.69 ± 6.14%) than 2D phase-contrast (28.77 ± 21.01%).
CONCLUSION: High-resolution 4D flow can assess hepatic artery anatomy and hemodynamics with improved accuracy, greater vessel visibility, better interobserver reliability, and internal consistency.
KEY POINTS: • Motion-suppressed Cartesian four-dimensional (4D) flow MRI with higher spatial resolution provides more accurate measurements even when accepted respiratory motion exceeds voxel size. • 4D flow MRI with higher spatial resolution provides substantial interobserver agreement for visualization of hepatic artery branches. • Lower peak and average velocities and a trend toward better internal consistency were observed with 4D flow MRI as compared to 2D phase-contrast.