第一代临床双源光子计数CT:超低剂量定量能谱成像
在体模上比较第一代双源光子计数CT(PCCT)与双源双能CT(DECT)在0.4-15.0 mGy剂量范围内的图像特征,PCCT在低剂量下表现更优,最低剂量时电子背景噪声较DECT 70/Sn150 kVp平均降低74%、较100/Sn150 kVp降低60%。
要点
- PCCT在超低剂量下仍可提供精确可靠的CT值,不受电子背景噪声干扰。
- 最低剂量时PCCT电子背景噪声较DECT 70/Sn150 kVp平均降低74%(p值约10⁻⁸),较100/Sn150 kVp降低60%(p值约10⁻⁶)。
- 两种能谱平台在宽辐射剂量范围内均表现良好,PCCT在低辐射剂量下更优。
核心结论
- PCCT在宽剂量范围内可提供可靠CT值且不受电子背景噪声干扰,诊断获益不仅在于超低剂量下的定量,也在于肥胖患者的成像。
研究设计与样本
- 体模研究:多能CT体模在带与不带延伸环条件下于两台扫描仪上成像,辐射剂量范围CTDIvol 0.4-15.0 mGy。
- PCCT采用120 kVp采集,DECT采用70/Sn150 kVp与100/Sn150 kVp采集。
- 使用多种组织插入物评估40-190 keV虚拟单能图像CT值的精确性与重复性。
诊断标准与关键数值
- PCCT在多种插入物和辐射暴露水平下CT值测量精确性高,两种扫描仪在宽辐射暴露范围内均表现良好,PCCT在低暴露时优于DECT。
- 最低辐射暴露时PCCT电子背景噪声较DECT 70/Sn150 kVp平均降低74%(p值约10⁻⁸),较DECT 100/Sn150 kVp平均降低60%(p值约10⁻⁶)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: Evaluation of image characteristics at ultra-low radiation dose levels of a first-generation dual-source photon-counting computed tomography (PCCT) compared to a dual-source dual-energy CT (DECT) scanner.
METHODS: A multi-energy CT phantom was imaged with and without an extension ring on both scanners over a range of radiation dose levels (CTDIvol 0.4-15.0 mGy). Scans were performed in different modes of acquisition for PCCT with 120 kVp and DECT with 70/Sn150 kVp and 100/Sn150 kVp. Various tissue inserts were used to characterize the precision and repeatability of Hounsfield units (HUs) on virtual mono-energetic images between 40 and 190 keV. Image noise was additionally investigated at an ultra-low radiation dose to illustrate PCCT's ability to remove electronic background noise.
RESULTS: Our results demonstrate the high precision of HU measurements for a wide range of inserts and radiation exposure levels with PCCT. We report high performance for both scanners across a wide range of radiation exposure levels, with PCCT outperforming at low exposures compared to DECT. PCCT scans at the lowest radiation exposures illustrate significant reduction in electronic background noise, with a mean percent reduction of 74% (p value ~ 10-8) compared to DECT 70/Sn150 kVp and 60% (p value ~ 10-6) compared to DECT 100/Sn150 kVp.
CONCLUSIONS: This paper reports the first experiences with a clinical dual-source PCCT. PCCT provides reliable HUs without disruption from electronic background noise for a wide range of dose values. Diagnostic benefits are not only for quantification at an ultra-low dose but also for imaging of obese patients.
KEY POINTS: PCCT scanners provide precise and reliable Hounsfield units at ultra-low dose levels. The influence of electronic background noise can be removed at ultra-low-dose acquisitions with PCCT. Both spectral platforms have high performance along a wide range of radiation exposure levels, with PCCT outperforming at low radiation exposures.