儿童腹盆腔CT深度学习重建算法与滤波反投影及迭代重建算法的比较
本研究回顾性纳入120例儿童腹盆腔增强CT,比较滤波反投影、混合迭代重建(ASiR-V 50%和100%)与深度学习重建(TrueFidelity低、中、高强度)的噪声特性、空间分辨率和主观图像质量。结果显示深度学习重建的噪声特性与空间分辨率优于混合迭代重建,高强度深度学习重建的整体图像质量主观评分优于ASiR-V。
要点
- 深度学习重建(TrueFidelity)在儿童腹盆腔CT中噪声特性与空间分辨率优于混合迭代重建(ASiR-V)。
- 高强度深度学习重建的整体图像质量主观评分优于ASiR-V(p < 0.001)。
- 深度学习重建的边缘上升距离短于ASiR-V和滤波反投影(p < 0.001)。
核心结论
- 对于儿童腹盆腔CT,深度学习重建算法可能提供优于混合迭代重建算法的噪声特性和更好的空间分辨率。
研究设计与样本
- 回顾性研究,纳入2020年5月至2020年10月间120例儿童患者(平均年龄±标准差8.7±5.2岁,60例男性)的腹盆腔增强CT扫描。
诊断标准与关键数值
- 噪声功率谱噪声幅度:ASiR-V 100%显著低于深度学习重建;噪声功率谱峰值:ASiR-V 100%显著高于高强度和中强度深度学习重建(p < 0.001)。
- 噪声功率谱平均空间频率:深度学习重建高于ASiR-V(p < 0.001)。
- 边缘上升距离:深度学习重建短于ASiR-V和滤波反投影(p < 0.001)。
- 定性视觉分析:高强度深度学习重建的整体图像质量优于ASiR-V(p < 0.001)。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To compare a deep learning-based reconstruction (DLR) algorithm for pediatric abdominopelvic computed tomography (CT) with filtered back projection (FBP) and iterative reconstruction (IR) algorithms.
MATERIALS AND METHODS: Post-contrast abdominopelvic CT scans obtained from 120 pediatric patients (mean age ± standard deviation, 8.7 ± 5.2 years; 60 males) between May 2020 and October 2020 were evaluated in this retrospective study. Images were reconstructed using FBP, a hybrid IR algorithm (ASiR-V) with blending factors of 50% and 100% (AV50 and AV100, respectively), and a DLR algorithm (TrueFidelity) with three strength levels (low, medium, and high). Noise power spectrum (NPS) and edge rise distance (ERD) were used to evaluate noise characteristics and spatial resolution, respectively. Image noise, edge definition, overall image quality, lesion detectability and conspicuity, and artifacts were qualitatively scored by two pediatric radiologists, and the scores of the two reviewers were averaged. A repeated-measures analysis of variance followed by the Bonferroni post-hoc test was used to compare NPS and ERD among the six reconstruction methods. The Friedman rank sum test followed by the Nemenyi-Wilcoxon-Wilcox all-pairs test was used to compare the results of the qualitative visual analysis among the six reconstruction methods.
RESULTS: The NPS noise magnitude of AV100 was significantly lower than that of the DLR, whereas the NPS peak of AV100 was significantly higher than that of the high- and medium-strength DLR (p < 0.001). The NPS average spatial frequencies were higher for DLR than for ASiR-V (p < 0.001). ERD was shorter with DLR than with ASiR-V and FBP (p < 0.001). Qualitative visual analysis revealed better overall image quality with high-strength DLR than with ASiR-V (p < 0.001).
CONCLUSION: For pediatric abdominopelvic CT, the DLR algorithm may provide improved noise characteristics and better spatial resolution than the hybrid IR algorithm.
引用
- Son, Kim, Hwang, et al. Comparison of a Deep Learning-Based Reconstruction Algorithm with Filtered Back Projection and Iterative Reconstruction Algorithms for Pediatric Abdominopelvic CT. Korean J Radiol. 2022. PMID: 35695313. doi:10.3348/kjr.2021.0466
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。