文献速递外周血管腹部技术与安全US

超声探头压力:腹主动脉瘤测量误差的一个未被探索的来源

Ultrasound Transducer Pressure: An Unexplored Source of Abdominal Aortic Aneurysm Measurement Error.
PubMed 文献 · Ultrasound Med Biol · 2022-06-09

横断面研究纳入47例腹主动脉瘤患者(前后径中位数46 mm,范围30–76 mm),以预定义厚度15 mm、质量150 g的凝胶垫估测探头压力。用力加压使前后径中位数从46.1 mm降至39.4 mm(p=0.001),轻压与重压间平均直径偏倚6.1 mm(95%CI 4.9–7.3),探头压力可通过凝胶垫形变测量。

要点

  • 用力加压可使腹主动脉瘤前后径中位数由46.1 mm降至39.4 mm,差异有统计学意义(p=0.001)。
  • 轻压与重压间平均直径偏倚为6.1 mm(95%CI 4.9–7.3)。
  • 低技术凝胶垫的形变可用于估测施加的探头压力。

核心结论

  • 用力加压探头可显著降低腹主动脉瘤前后径测量值,轻压与重压间平均直径偏倚为6.1 mm(95%CI 4.9–7.3)。

研究设计与样本

  • 横断面研究,纳入47例腹主动脉瘤患者,前后径中位数46 mm(范围30–76 mm),33/47(70%)体重指数>25;以预定义厚度15 mm、质量150 g的超声兼容凝胶垫置于探头与受检者之间估测探头压力。

诊断标准与关键数值

  • 用力加压探头使前后径中位数从46.1 mm(范围29.7–76.3)降至39.4 mm(范围21.7–67.5),p=0.001。
  • 轻压与重压间平均直径偏倚为6.1 mm(95%CI 4.9–7.3)。
  • 探头压力可通过置于探头与受检者之间的低技术凝胶垫的形变换算测得,且施加压力存在显著变异。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

This study was aimed at quantifying the abdominal aortic aneurysm (AAA) compression phenomenon and assessing the use of a new tool to estimate transducer pressure in vivo. In this cross-sectional study, 47 participants with AAA and a median anterior-to-posterior (AP) AAA diameter of 46 mm (range: 30-76 mm) were included. The majority of the patients were overweight with body mass indexes >25 (33/47, 70%). A standardized ultrasound (US)-compatible gel pad, with a pre-defined thickness (15 mm) and mass (150 g), was interposed between the US transducer and participant to estimate the applied transducer pressure. A firm transducer pressure significantly lowered the median AP diameter from 46.1 mm (range: 29.7-76.3) to 39.4 mm (range: 21.7-67.5) (p = 0.001). The mean diameter bias between light and firm transducer pressure was 6.1 mm (95% confidence interval: 4.9-7.3). The applied transducer pressure varied significantly and could be measured by translating the deformation of a low-tech gel pad interposed between the US transducer and the participant.

来源:Ultrasound Med Biol,查看原文。版权归原作者所有。

引用

  1. Ghulam, Svendsen, Zielinski, et al. Ultrasound Transducer Pressure: An Unexplored Source of Abdominal Aortic Aneurysm Measurement Error. Ultrasound Med Biol. 2022. PMID: 35691732. doi:10.1016/j.ultrasmedbio.2022.05.003

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。