COVID-19疫情期间远程超声机器人的应用:可行性研究
研究纳入雷神山医院22例新型冠状病毒肺炎患者,由远程超声机器人(Imabot公司,武汉)与常规超声分别扫描甲状腺、颈部血管、肝胆和肾脏,比较两种方法的超声诊断。110个部位共44个阳性病灶,机器人法检出40个、漏诊4个、误诊1个,常规法检出44个、漏诊1个,两种方法准确率差异无统计学意义(P>.05)。结论认为远程超声机器人用于COVID-19患者常见腹部、血管、浅表器官病变诊断是可行的。
要点
- 22例COVID-19患者110个部位共44个阳性病灶,机器人法检出40个,常规法检出44个。
- 机器人法漏诊4个(胆囊小息肉2个、肝小血管瘤1个、肾小囊肿1个),误诊1个(正常颈动脉误判为粥样硬化斑块)。
- 机器人法与常规法准确率差异无统计学意义(P>.05)。
核心结论
- 远程超声机器人应用于COVID-19疫情期间可满足患者诊疗标准,与常规超声相比能以可接受的准确率提供足够的超声信息以诊断常见腹部、血管、浅表器官病变,该方法可行。
研究设计与样本
- 可行性研究,纳入雷神山医院22例新型冠状病毒肺炎患者,由Imabot公司(武汉)制造的远程超声机器人与常规方法分别扫描甲状腺、颈部血管、肝胆和肾脏,两种方法获得的超声图像混合后由一名经验丰富的超声放射科医生双盲诊断。
诊断标准与关键数值
- 22例患者110个部位共44个阳性病灶:机器人法检出40个,漏诊4个(胆囊小息肉2个、肝小血管瘤1个、肾小囊肿1个),误诊1个(正常颈动脉误诊为颈动脉粥样硬化斑块)。
- 常规法检出44个阳性病灶,漏诊1个(肝小囊肿)。
- 机器人法与常规法准确率比较采用四格表卡方检验,差异无统计学意义(P>.05)。
临床应用与阅片要点
- 作者认为远程超声机器人应用于COVID-19疫情期间可满足患者诊疗标准,该方法可行,与常规超声相比能以可接受的准确率提供诊断常见腹部、血管、浅表器官病变所需的超声信息。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: To investigate the accuracy of ultrasonic diagnosis using the tele-ultrasound robot in Leishen Shan Hospital.
METHOD: Twenty-two patients with novel coronavirus pneumonia from Leishen Shan Hospital voluntarily participated in this study. Their thyroids, neck vessels, hepatobiliaries and kidneys were scanned by both a tele-ultrasound robot manufactured by Imabot Co., Ltd, Wuhan and conventional method. The ultrasound diagnosis of each patient was compared, and the ultrasound images obtained by the two methods were mixed together and double-blindly diagnosed by an experienced ultrasound radiologist.
RESULTS: There were 44 positive lesions in 110 sites of 22 patients. Of which the two methods, 40 positive lesions were detected by the robotic method with 4 lesions missed (2 small polyps of gallbladder, 1 small hemangioma of liver and 1 small cyst of kidney) and 1 lesion misdiagnosed (normal carotid artery was misdiagnosed as carotid atherosclerotic plaque); 44 positive lesions were detected by conventional method with 1 small cyst of the liver was missed. There was no statistically significant difference in the accuracy rate between the robotic method and the conventional method using the chi-square test of the four-grid data (P>.05).
CONCLUSION: The application of tele-ultrasound robot meets the standard of patient care during the pandemic. The method is feasible to provide adequate ultrasound information to diagnose common abdominal, vascular, superficial organ pathologies in patients with COVID-19 with acceptable accuracy compared with a conventional ultrasound scan.