新型三维容积超声方法用于准确量化动脉粥样硬化斑块体积
本研究在猪颈动脉和股动脉标本中验证了新型电子扫查三维矩阵探头及配套三维斑块定量软件测量斑块体积的准确性,并与组织学金标准对照;随后在20例中高危个体中比较该探头与既往机械扫查三维探头及二维定量软件的结果。离体ICC为0.992,临床ICC为0.997,三维软件较二维多层面法分析时间平均缩短46%。
要点
- 新型电子扫查三维矩阵探头测得的斑块体积与组织学金标准高度一致(ICC 0.992)。
- 临床中电子扫查与机械扫查三维超声测值高度一致(ICC 0.997)。
- 三维斑块定量软件较二维多层面法分析时间平均缩短46%。
核心结论
- 采用新型矩阵探头技术及改进的三维斑块定量软件进行三维血管超声,可简化颈动脉和股动脉早期(小)及中晚期斑块的准确体积测量。
研究设计与样本
- 离体研究在猪颈动脉和股动脉标本中测量19个动脉粥样硬化斑块体积,以组织学为金标准;临床可行性研究纳入20例中高危个体(平均年龄74.6±4.45岁,40%为男性),比较电子扫查三维矩阵探头与既往验证的机械扫查三维探头及二维斑块定量软件。
诊断标准与关键数值
- 离体研究中三维矩阵探头测值与组织学金标准的组内相关系数ICC为0.992(95%CI 0.978-0.997);临床研究中电子扫查与机械扫查三维超声测值的ICC为0.997(95%CI 0.995-0.998);二维与三维斑块定量软件的ICC为0.999(95%CI 0.998-0.999)。
临床应用与阅片要点
- 离体斑块体积范围为0.76 µL-56.30 μL;临床80个扫描区域中64个(36个颈动脉、28个股动脉)存在动脉粥样硬化,测得斑块体积为10 μL-859 μL;三维软件较二维多层面法分析时间平均缩短46%。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
BACKGROUND: Carotid and femoral plaque burden is a recognized biomarker of cardiovascular disease risk. A new electronic-sweep 3-dimensional (3D)-matrix transducer method can improve the functionality and image quality of vascular ultrasound atherosclerosis imaging.
OBJECTIVES: This study aimed to validate this method for plaque volume measurement in early and intermediate-advanced plaques in the carotid and femoral territories.
METHODS: Plaque volumes were measured ex vivo in pig carotid and femoral artery specimens by 3-dimensional vascular ultrasound (3DVUS) using a 3D-matrix (electronic-sweep) transducer and its associated 3D plaque quantification software, and were compared with gold-standard histology. To test the clinical feasibility and accuracy of the 3D-matrix transducer, an experiment was conducted in intermediate-high risk individuals with carotid and femoral atherosclerosis. The results were compared with those obtained using the previously validated mechanical-sweep 3D transducer and established 2-dimensional (2D)-based plaque quantification software.
RESULTS: In the ex vivo study, the authors assessed 19 atherosclerotic plaques (plaque volume, 0.76 µL-56.30 μL), finding strong agreement between measurements with the 3D-matrix transducer and the histological gold-standard (intraclass correlation coefficient [ICC]: 0.992; [95% CI: 0.978-0.997]). In the clinical analysis of 20 patients (mean age 74.6 ± 4.45 years; 40% men), the authors found 64 (36 carotid and 28 femoral) of 80 scanned territories with atherosclerosis (measured atherosclerotic volume, 10 μL-859 μL). There was strong agreement between measurements made from electronic-sweep and mechanical-sweep 3DVUS transducers (ICC: 0.997 [95% CI: 0.995-0.998]). Agreement was also high between plaque volumes estimated by the 2D and 3D plaque quantification software applications (ICC: 0.999 [95% CI: 0.998-0.999]). Analysis time was significantly shorter with the 3D plaque quantification software than with the 2D multislice approach with a mean time reduction of 46%.
CONCLUSIONS: 3DVUS using new matrix transducer technology, together with improved 3D plaque quantification software, simplifies the accurate volume measurement of early (small) and intermediate-advanced plaques located in carotid and femoral arteries.