文献速递腹部技术与安全USMR

超声定量评估肝脂肪变性:以MRI质子密度脂肪分数为参考标准的前瞻性比较研究

Quantification of Hepatic Steatosis by Ultrasound: Prospective Comparison With MRI Proton Density Fat Fraction as Reference Standard.
PubMed 文献 · AJR Am J Roentgenol · 2022-06-08

这项前瞻性横断面研究纳入56名超重和肥胖的青少年及成人,在同一次就诊中接受超声(深腹部探头)和MRI肝脏检查,比较超声衍生脂肪分数(UDFF)与MRI质子密度脂肪分数(PDFF)的一致性。UDFFoverall与MRI PDFF呈强正相关(ρ=0.82,P<.001),组内相关系数0.84(95%CI 0.59-0.93),平均偏倚4.0%(95%一致性界限-7.9%至15.9%)。UDFFoverall诊断MRI PDFF≥5.5%的AUC为0.90(95%CI 0.79-0.96), cutoff>5%时敏感度94.1%、特异度63.6%。

要点

  • UDFF与MRI PDFF强相关(ρ=0.82)且一致性良好(ICC 0.84)。
  • UDFFoverall诊断MRI PDFF≥5.5%的AUC为0.90,cutoff>5%时敏感度94.1%、特异度63.6%。
  • 减少单次测量次数(前5次、前3次或第1次)与总体中位数偏倚相似,提示减少测量次数可能足够。

核心结论

  • UDFF测量肝脂肪变性与MRI PDFF测量结果高度一致;UDFFoverall cutoff>5%检测MRI PDFF≥5.5%具有高AUC和高敏感度,但特异度低。

研究设计与样本

  • 前瞻性横断面研究,纳入56名年龄≥16岁的超重和肥胖青少年及成人,于2020年8月至2020年10月单次就诊期间接受研究性超声(深腹部探头)和MRI肝脏检查;超声包括三次UDFF采集、每次五个测量(共15个测量),计算总体中位数(UDFFoverall);MRI包括三次PDFF采集并计算总体中位PDFF;以MRI PDFF为参考标准。

诊断标准与关键数值

  • UDFFoverall诊断MRI PDFF≥5.5%的AUC为0.90(95%CI 0.79-0.96);UDFFoverall cutoff>5%时敏感度94.1%、特异度63.6%;UDFFoverall与MRI PDFF的Spearman相关系数ρ=0.82(P<.001),组内相关系数0.84(95%CI 0.59-0.93);平均偏倚4.0%(95%一致性界限-7.9%至15.9%),用前五次测量、前三次测量或第一次测量汇总UDFF时偏倚分别为4.4%、4.4%、4.6%。

临床应用与阅片要点

  • UDFF可能在检测肝脂肪变性中具有临床作用;减少单次测量次数可能足以确定总体UDFF值。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

BACKGROUND. Multiple ultrasound platforms now provide quantitative measures of hepatic steatosis. One such measure is the ultrasound-derived fat fraction (UDFF), which combines attenuation and backscatter quantification. OBJECTIVE. The purpose of this study was to characterize agreement between UDFF and MRI proton-density fat fraction (PDFF) measurements. METHODS. This prospective cross-sectional study enrolled 56 overweight and obese adolescents and adults (age ≥ 16 years) who underwent investigational ultrasound (deep abdominal transducer) and MRI examinations of the liver during a single visit from August 2020 to October 2020. Ultrasound examinations included three UDFF acquisitions of five measurements each (15 measurements total), and an overall median of medians was computed (UDFFoverall). MRI examinations included three PDFF acquisitions with calculation of an overall median PDFF. Spearman rank-order correlation was computed between UDFF and MRI PDFF measurements. Intraclass correlation coefficients and Bland-Altman difference plots were used to assess agreement. ROC curves were used to assess diagnostic performance of UDFF for detecting MRI PDFF of 5.5% or more. RESULTS. Median participant age was 32.5 years (IQR, 24.0-39.0 years); 40 participants were female, and 16 were male. A total of 34 (60.7%) participants had an MRI PDFF of 5.5% or more. UDFFoverall was 10.5% (IQR, 5.0-20.0%); median MRI PDFF was 6.1% (IQR, 3.4-13.7%). UDFFoverall was positively associated with MRI PDFF (ρ, 0.82; p < .001; intraclass correlation coefficient, 0.84 [95% CI, 0.59-0.93]). Mean bias between UDFF and PDFF was 4.0% (95% limits of agreement, -7.9% to 15.9%), with similar bias if summarizing UDFF by the first five measurements (4.4%), first three measurements (4.4%), or first measurement (4.6%). UDFFoverall AUC was 0.90 (95% CI, 0.79-0.96) for MRI PDFF of 5.5% or more; AUC was not significantly different when it was based on the number of UDFF measurements (p = .11-.97 for all pairwise AUC comparisons). UDFFoverall cutoff of more than 5% had sensitivity of 94.1% and specificity of 63.6% for diagnosing MRI PDFF of 5.5% or more. CONCLUSION. Measurements of hepatic steatosis using UDFF show strong agreement with measurements by MRI PDFF. A UDFFoverall cutoff of more than 5% provides high AUC and sensitivity, albeit low specificity, for detection of MRI PDFF of 5.5% or more. CLINICAL IMPACT. UDFF may have a clinical role in detection of hepatic steatosis. A reduced number of individual measurements is likely sufficient for determining an overall UDFF value. TRIAL REGISTRATION. ClinicalTrials.gov: NCT04523584.

来源:AJR Am J Roentgenol,查看原文。版权归原作者所有。

引用

  1. Dillman, Thapaliya, Tkach, et al. Quantification of Hepatic Steatosis by Ultrasound: Prospective Comparison With MRI Proton Density Fat Fraction as Reference Standard. AJR Am J Roentgenol. 2022. PMID: 35674351. doi:10.2214/AJR.22.27878

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。