采用深度学习重建与新型滤线板的新型320排CT行CT透视引导介入的剂量长度乘积与结局
回顾性比较Prism与Vision两台320排CT行CT透视引导介入的剂量长度乘积与结局。Prism组总DLP显著低于Vision组(活检278 vs 548 mGy*cm,引流抽吸246 vs 667 mGy*cm),技术成功率、活检临床成功率及并发症无显著差异。
要点
- Prism组活检总DLP 278 vs Vision组548 mGy*cm,降低49%
- Prism组引流抽吸总DLP 246 vs Vision组667 mGy*cm,降低63%
- 两组技术成功率、活检临床成功率及并发症无显著差异
核心结论
- 在Prism上行CT透视引导介入可降低总DLP而不降低介入操作性能。
研究设计与样本
- 回顾性研究,Prism组85例、Vision组87例CT透视引导介入,比较技术成功率、活检临床成功率、并发症及总DLP,时间范围为2019年2月至2021年1月。
诊断标准与关键数值
- 技术成功率Prism组98.8% vs Vision组96.6%,p=0.621
- 活检临床成功率Prism组93.4% vs Vision组92.9%,p=1.000
- 轻微并发症Prism组7.1% vs Vision组8.0%,p=0.807
- 严重并发症Prism组3.5% vs Vision组0%,p=0.119
- 活检总DLP Prism组278 vs Vision组548 mGy*cm,p<0.001,Prism组降低49%
临床应用与阅片要点
- 使用Prism行CT透视引导介入时,活检和引流抽吸的总DLP均较Vision降低,可减少辐射剂量。
- Prism组与Vision组的技术成功率、活检临床成功率及并发症无显著差异。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: To investigate the dose length product (DLP) and outcomes of CT fluoroscopy (CTF)-guided interventions using a novel 320-detector row CT scanner with deep-learning reconstruction (DLR) and a new bow-tie filter (i.e., Aquilion ONE Prism Edition) and compare with a 320-detector row CT system without DLR and the new bow-tie filter (i.e., Aquilion ONE Vision Edition) (Vision).
METHODS: CTF-guided interventions performed using Prism and Vision were retrospectively investigated in terms of the technical success rates, clinical success rates of biopsies, complications, DLPs of total CT scans (total DLPs) from February 2019 to January 2021. The total CT scans included pre-interventional CT scans, CTF scans during the CTF-guided procedure, additional CT scans for additional treatment, CTF scans for additional treatment, and post-interventional CT scans.
RESULTS: In this study, 87 and 85 CTF-guided interventions were performed using Vision (Vision group) and Prism (Prism group), respectively. There was no significant difference in the technical success rate (96.6% vs 98.8%, p = 0.621), clinical success rate of biopsies (92.9% vs 93.4%, p = 1.000), and minor (8.0% vs 7.1%, p = 0.807) and major (0% vs 3.5%, p = 0.119) complications between the Prism and Vision groups. The total DLPs for the Prism group were significantly lower than those for the Vision group regardless of the procedure (278 vs 548 mGy*cm, p < 0.001, in the biopsy and 246 vs 667 mGy*cm, p < 0.001, in the drainage and aspiration).
CONCLUSIONS: CTF-guided interventions on Prism reduce the total DLP without performance degradation of the intervention.
ADVANCES IN KNOWLEDGE: The total DLPs of biopsies and drainages/aspirations in the Prism group decreased by 49 and 63%, respectively.
引用
- Matsumoto, Endo, Yamamoto, et al. Dose length product and outcome of CT fluoroscopy-guided interventions using a new 320-detector row CT scanner with deep-learning reconstruction and new bow-tie filter. Br J Radiol. 2022. PMID: 35671099. doi:10.1259/bjr.20211159
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。