读出分段平面回波成像联合二维导航校正(RESOLVE):一种改善前列腺扩散加权成像图像质量的替代序列
这项单中心前瞻性研究纳入162例疑似前列腺癌男性,比较RESOLVE与单次激发平面回波成像(ss-EPI)采集前列腺扩散加权成像的图像质量;RESOLVE在几何畸变、病灶显示和解剖细节显示上均显著优于ss-EPI,几何畸变评分>3者仅见于ss-EPI(12.3%)。
要点
- RESOLVE序列前列腺DWI几何畸变平均评分1.16,低于ss-EPI的1.61(p<.01)。
- ss-EPI图像中12.3%(20/162)几何畸变评分>3,RESOLVE对应图像均<3分(p<.001)。
- RESOLVE病灶显示评分1.35优于ss-EPI的1.53(p<.002),解剖细节2.60优于2.68(p<.001)。
核心结论
- 与常规ss-EPI相比,采用RESOLVE采集前列腺DWI可显著改善图像质量并减少几何畸变;RESOLVE可作为ss-EPI的替代或替换序列用于前列腺DWI采集。
研究设计与样本
- 单中心前瞻性研究队列,纳入162例疑似前列腺癌男性,行3特斯拉多参数MRI(3T-mpMRI);两名腹部放射科医师在不知临床信息的情况下,分别以5分制(1=优,5=差)先对ss-EPI-DWI、后对RESOLVE-DWI独立评估几何畸变、直肠扩张程度、病灶显示和解剖细节显示,取两名读者评分的平均值作为最终评分。
关键影像表现/征象
- ss-EPI图像中12.3%(20/162)几何畸变评分>3,而RESOLVE所有图像几何畸变评分均<3(p<.001)。
- RESOLVE平均图像畸变评分1.16,低于ss-EPI的1.61(p<.01,无论直肠气体如何;按直肠扩张程度分层时p<.05)。
- RESOLVE病灶显示评分1.35,优于ss-EPI的1.53(p<.002)。
- RESOLVE前列腺解剖细节显示评分2.60,优于ss-EPI的2.68(p<.001)。
诊断标准与关键数值
- 图像质量采用5分制评分:1=优,5=差;两名读者评分的平均值作为最终评分。
- ss-EPI与RESOLVE的图像质量参数评分均显示良好至优秀的一致性(good-to-excellent interreader agreement)。
临床应用与阅片要点
- RESOLVE可作为前列腺DWI的替代或替换序列,以减少几何畸变并改善病灶显示和解剖细节显示。
- ss-EPI图像中几何畸变评分>3者占12.3%(20/162),采用RESOLVE后对应图像几何畸变评分均<3。
- 直肠气体存在时RESOLVE畸变评分仍低于ss-EPI(p<.01),按直肠扩张程度分层亦如此(p<.05)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVE: We aimed to investigate if the use of read-out segmented echoplanar imaging with additional two-dimensional navigator correction (Readout Segmentation of Long Variable Echo, RESOLVE) for acquiring prostate diffusion-weighted imaging (DWI) improves image quality, compared to single-shot echoplanar imaging (ss-EPI).
METHODS: This single-center prospective study cohort included 162 males with suspected prostate cancer, who underwent 3 Tesla multiparametric MRI (3T-mpMRI). Two abdominal radiologists, blinded to the clinical information, separately reviewed each 3T-mpMRI study to rank geometrical distortion, degree of rectal distention, lesion conspicuity, and anatomic details delineation first on ss-EPI-DWI and later on RESOLVE-DWI using 5-point scales (1 = excellent, 5 = poor). The average of the ranking scores given by two readers was generated and used as the final score.
RESULTS: There was good-to-excellent interreader agreement for scoring image quality parameters on both ss-EPI and RESOLVE. Geometrical distortion scores > 3 was seen in 12.3% (20/162) of ss-EPI images, with all having geometrical distortion score <3 on RESOLVE (p < .001). The mean image distortion score was significantly less on RESOLVE than ss-EPI (1.16 vs 1.61, p < .01 regardless of rectal gas, p< .05 when stratified by the degree of rectal distention ). RESOLVE was superior to ss-EPI for lesion conspicuity (mean 1.35 vs 1.53, p< .002) and anatomic delineation (2.60 vs 2.68, p< .001) of prostate on DWI.
CONCLUSION: Compared to conventional ss-EPI, the use of RESOLVE for acquisition of prostate DWI resulted in significantly enhanced image quality and reduced geometrical distortion.
ADVANCES IN KNOWLEDGE: RESOLVE could be an alternative or replacement of ss-EPI for acquiring prostate DWI with significantly less geometrical distortion and significantly improved lesion conspicuity and anatomic delineation.