光子计数探测器CT的腹部虚拟平扫成像

Virtual Noncontrast Abdominal Imaging with Photon-counting Detector CT.
PubMed 文献 · Radiology · 2022-06-07

这项回顾性研究纳入100例接受光子计数探测器CT三期检查的成年患者,评估动脉期和门静脉期重建的虚拟平扫图像与真实平扫相比的衰减误差和图像质量。虚拟平扫的CT衰减值准确,76%的测量误差小于5 HU、95%小于10 HU,诊断质量在两位读者中分别达到99%和100%,但主观图像质量评分在各分类均低于真实平扫。

要点

  • 虚拟平扫CT衰减误差小于5 HU占76%、小于10 HU占95%,动脉期与门静脉期重建误差无显著差异。
  • 两位读者对虚拟平扫的诊断质量分别达99%和100%,但主观图像质量各分类评分均低于真实平扫(P < .001)。
  • 体模中虚拟平扫噪声高33%、噪声纹理更斑驳,空间分辨率相似,可检测性指数大于20。

核心结论

  • 光子计数探测器CT动脉期和门静脉期重建的腹部虚拟平扫图像与真实平扫相比,CT衰减准确且图像质量良好。

研究设计与样本

  • 回顾性研究,纳入2021年7月至2021年10月连续接受光子计数探测器CT三期检查的成年患者100例(平均年龄72岁±10,81例男性),由两名读者测量衰减误差并盲法评估图像质量,另进行体模评估。

关键影像表现/征象

  • 虚拟平扫图像衰减值在两位读者间一致性为α = .60。
  • 虚拟平扫衰减误差小于5 HU占76%的测量,小于10 HU占95%的测量。
  • 动脉期与门静脉期重建的虚拟平扫误差分别为3.3 HU与3.5 HU,差异无统计学意义(P = .16)。
  • 体模中虚拟平扫较真实平扫噪声高33%、噪声纹理更斑驳、空间分辨率相似,图像质量较差但总体良好(可检测性指数>20)。

诊断标准与关键数值

  • 主观图像质量采用五点离散视觉量表(5 = 优秀,1 = 不可诊断),总体图像质量大于或等于3判定为可诊断。
  • 虚拟平扫诊断质量达到可诊断的比例:读者1为99%,读者2为100%。
  • 体模可检测性指数大于或等于5提示高诊断准确性,虚拟平扫可检测性指数>20。

临床应用与阅片要点

  • 虚拟平扫图像主观图像质量在所有评估分类中均低于真实平扫(均P < .001),阅片时需注意其噪声和噪声纹理与真实平扫的差异。
  • 动脉期和门静脉期重建的虚拟平扫衰减误差无显著差异(3.3 HU vs 3.5 HU,P = .16),两期均可用于生成虚拟平扫。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

Background Accurate CT attenuation and diagnostic quality of virtual noncontrast (VNC) images acquired with photon-counting detector (PCD) CT are needed to replace true noncontrast (TNC) scans. Purpose To assess the attenuation errors and image quality of VNC images from abdominal PCD CT compared with TNC images. Materials and Methods In this retrospective study, consecutive adult patients who underwent a triphasic examination with PCD CT from July 2021 to October 2021 were included. VNC images were reconstructed from arterial and portal venous phase CT. The absolute attenuation error of VNC compared with TNC images was measured in multiple structures by two readers. Then, two readers blinded to image reconstruction assessed the overall image quality, image noise, noise texture, and delineation of small structures using five-point discrete visual scales (5 = excellent, 1 = nondiagnostic). Overall image quality greater than or equal to 3 was deemed diagnostic. In a phantom, noise texture, spatial resolution, and detectability index were assessed. A detectability index greater than or equal to 5 indicated high diagnostic accuracy. Interreader agreement was evaluated using the Krippendorff α coefficient. The paired t test and Friedman test were applied to compare objective and subjective results. Results Overall, 100 patients (mean age, 72 years ± 10 [SD]; 81 men) were included. In patients, VNC image attenuation values were consistent between readers (α = .60), with errors less than 5 HU in 76% and less than 10 HU in 95% of measurements. There was no evidence of a difference in error of VNC images from arterial or portal venous phase CT (3.3 HU vs 3.5 HU, P = .16). Subjective image quality was rated lower in VNC images for all categories (all, P < .001). Diagnostic quality of VNC images was reached in 99% and 100% of patients for readers 1 and 2, respectively. In the phantom, VNC images exhibited 33% higher noise, blotchier noise texture, similar spatial resolution, and inferior but overall good image quality (detectability index >20) compared with TNC images. Conclusion Abdominal virtual noncontrast images from the arterial and portal venous phase of photon-counting detector CT yielded accurate CT attenuation and good image quality compared with true noncontrast images. © RSNA, 2022 Online supplemental material is available for this article See also the editorial by Sosna in this issue.