COVID-19疫情期间佩戴的FFP3呼吸口罩的MRI安全性、成像伪影和网格畸变评估

MRI safety, imaging artefacts, and grid distortion evaluated for FFP3 respiratory masks worn throughout the COVID-19 pandemic.
PubMed 文献 · Clin Radiol · 2022-05-16

研究评估8款商用FFP3口罩在3T MRI下的安全性,使用3D头模和3个临床序列测试伪影、网格畸变和局部加热。8款中5款含铁磁组件为MRI不安全,1款为MRI条件安全,2款对工作人员和患者均MRI安全;铁磁口罩前部最大网格畸变5mm,鼻梁处温度条未见局部加热。

要点

  • 8款FFP3口罩中5款含铁磁组件,属MRI不安全
  • 仅2款对MRI工作人员和患者均MRI安全,1款为MRI条件安全
  • 铁磁口罩前部头模最大网格畸变5mm,鼻梁处未见局部加热

核心结论

  • 研究证明评估FFP3呼吸口罩在MRI设备内及周围使用安全性的重要性,未来涉及FFP3口罩的研究可按本研究所述流程安全开展。

研究设计与样本

  • 在Cardiff大学脑影像研究中心使用3T Siemens Magnetom Prisma和64通道头颈线圈,对8款商用FFP3口罩用3个临床MRI序列在定制3D头模上评估成像伪影、网格畸变和局部加热。

关键影像表现/征象

  • 8款FFP3口罩中5款含铁磁组件,被认定为MRI不安全。
  • 1款口罩被认为MRI条件安全,仅2款对MRI工作人员和患者均认定为MRI安全。
  • 铁磁FFP3口罩在头模头部前部出现最大网格畸变5mm。
  • 置于头模鼻梁处的温度条未显示局部加热。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

AIM: To determine which filtering face piece (FFP3) respirators worn throughout the COVID-19 pandemic are safe for magnetic resonance imaging (MRI).

MATERIALS AND METHODS: Three clinical MRI sequences were performed to assess imaging artefacts, grid distortion, and local heating for eight commercially available FFP3 respirators. All examinations were performed at Cardiff University Brain Research Imaging Centre using a 3 T Siemens Magnetom Prisma with a 64-channel head and neck coil. Each FFP3 mask was positioned on a custom-developed three-dimensional (3D) head phantom for testing.

RESULTS: Five of the eight FFP3 masks contained ferromagnetic components and were regarded as "MRI unsafe". One mask was considered "MRI conditional" and only two masks were deemed "MRI safe" for both MRI staff and patients. Temperature strips positioned at the nasal bridge of the phantom did not exhibit local heating. A maximum grid distortion of 5 mm was seen in the anterior portion of the head of the ferromagnetic FFP3 masks.

CONCLUSION: This study has demonstrated the importance of assessing respiratory FFP3 masks for use in and around MRI machines. Future research involving FFP3 masks can be conducted safely by following the procedures laid out in this study.