文献速递心血管CT

光子计数探测器CT超高清冠状动脉CTA:可行性与图像特征分析

Ultra-High-Resolution Coronary CT Angiography With Photon-Counting Detector CT: Feasibility and Image Characterization.
PubMed 文献 · Invest Radiol · 2022-05-31

本研究纳入20例高冠状动脉钙化负荷患者,采用双源光子计数探测器CT超高清模式行冠状动脉CTA,评估8种血管核(Bv40至Bv89)结合量子迭代重建的图像质量,并以Bv40核、0.6 mm层厚为参考。超高清CCTA在所有患者中可行,Bv64和Bv72核在斑块及相邻管腔 delineation 上获读者最优评分(中位5分),Bv64核结合200×200 mm²视野和512×512矩阵可获得最佳图像质量。

要点

  • 超高清PCD-CCTA在所有患者中可行,中位钙化积分为479。
  • Bv64和Bv72核在斑块及相邻管腔勾画上获最优主观评分(中位5分)。
  • 随核锐度增加,SNR和CNR持续下降,图像噪声和血管锐度增加。

核心结论

  • 超高清PCD-CCTA可行,能显示钙化冠状动脉,图像质量优异、锐度高且 blooming 伪影减少;使用Bv64核、200×200 mm²视野和512×512矩阵可获得最佳质量的斑块特征及相邻管腔勾画。

研究设计与样本

  • 经机构审查委员会批准的研究,纳入20例(6例女性;平均年龄79±10岁;平均体重指数25.6±4.3 kg/m²)行PCD-CCTA超高清模式的患者;采用心电图门控双源螺旋模式,管电压120 kV,准直120×0.2 mm;视野200×200 mm²或150×150 mm²,矩阵512×512或1024×1024像素;使用8种锐度水平的血管核(Bv40、Bv44、Bv56、Bv60、Bv64、Bv72、Bv80、Bv89),量子迭代重建强度4,层厚0.2 mm;以Bv40核、量子迭代重建强度4、层厚0.6 mm为参考;体模扫描用于计算噪声功率谱。

关键影像表现/征象

  • 最大空间频率(f peak)在Bv40至Bv64核重建中逐渐向更高值偏移(0.15至0.56 mm⁻¹),但在Bv72至Bv89核重建中未见此偏移。
  • Bv40核、0.6 mm层厚重建显示最大blooming伪影(55.2%±9.8%)和最低血管锐度(477.1±73.6 ΔHU/mm),同时具有最高SNR(27.4±5.6)、最高CNR(32.9±6.6)和最低噪声(17.1±2.2 HU)。
  • 在0.2 mm层厚重建中,图像噪声、SNR、CNR、血管锐度和blooming伪影在各核间显著不同(所有P<0.001);随核锐度增高,SNR和CNR持续下降,图像噪声和血管锐度增加,Bv89核锐度最高(2383.4±787.1 ΔHU/mm)。
  • Blooming伪影从Bv40核(0.2 mm层厚;52.8%±9.2%)至Bv72核(39.7%±9.1%)持续下降。
  • 主观噪声感知与客观测量一致;斑块及相邻管腔勾画方面,Bv64和Bv72核(两者中位评分均为5分)获读者青睐,提供优异的斑块特征和管腔解剖勾画。

诊断标准与关键数值

  • 主观图像质量采用5分离散视觉量表评估图像噪声及冠状动脉斑块和相邻血管管腔的勾画。
  • Bv64和Bv72核在斑块及相邻管腔勾画上中位评分均为5分。
  • 参考重建为Bv40核、量子迭代重建强度4、层厚0.6 mm。

临床应用与阅片要点

  • 作者结论指出,使用Bv64核、200×200 mm²视野和512×512矩阵可获得最佳质量的冠状动脉斑块特征及相邻管腔勾画。
  • 随核锐度增加,SNR和CNR下降而血管锐度增加,阅片时需结合诊断目的选择重建核。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: The aim of this study was to evaluate the feasibility and quality of ultra-high-resolution coronary computed tomography angiography (CCTA) with dual-source photon-counting detector CT (PCD-CT) in patients with a high coronary calcium load, including an analysis of the optimal reconstruction kernel and matrix size.

MATERIALS AND METHODS: In this institutional review board-approved study, 20 patients (6 women; mean age, 79 ± 10 years; mean body mass index, 25.6 ± 4.3 kg/m 2 ) undergoing PCD-CCTA in the ultra-high-resolution mode were included. Ultra-high-resolution CCTA was acquired in an electrocardiography-gated dual-source spiral mode at a tube voltage of 120 kV and collimation of 120 × 0.2 mm. The field of view (FOV) and matrix sizes were adjusted to the resolution properties of the individual reconstruction kernels using a FOV of 200 × 200 mm 2 or 150 × 150 mm 2 and a matrix size of 512 × 512 pixels or 1024 × 1024 pixels, respectively. Images were reconstructed using vascular kernels of 8 sharpness levels (Bv40, Bv44, Bv56, Bv60, Bv64, Bv72, Bv80, and Bv89), using quantum iterative reconstruction (QIR) at a strength level of 4, and a slice thickness of 0.2 mm. Images with the Bv40 kernel, QIR at a strength level of 4, and a slice thickness of 0.6 mm served as the reference. Image noise, signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), vessel sharpness, and blooming artifacts were quantified. For subjective image quality, 2 blinded readers evaluated image noise and delineation of coronary artery plaques and the adjacent vessel lumen using a 5-point discrete visual scale. A phantom scan served to characterize image noise texture by calculating the noise power spectrum for every reconstruction kernel.

RESULTS: Maximum spatial frequency (f peak ) gradually shifted to higher values for reconstructions with the Bv40 to Bv64 kernel (0.15 to 0.56 mm -1 ), but not for reconstructions with the Bv72 to Bv89 kernel. Ultra-high-resolution CCTA was feasible in all patients (median calcium score, 479). In patients, reconstructions with the Bv40 kernel and a slice thickness of 0.6 mm showed largest blooming artifacts (55.2% ± 9.8%) and lowest vessel sharpness (477.1 ± 73.6 ΔHU/mm) while achieving highest SNR (27.4 ± 5.6) and CNR (32.9 ± 6.6) and lowest noise (17.1 ± 2.2 HU). Considering reconstructions with a slice thickness of 0.2 mm, image noise, SNR, CNR, vessel sharpness, and blooming artifacts significantly differed across kernels (all P 's < 0.001). With higher kernel sharpness, SNR and CNR continuously decreased, whereas image noise and vessel sharpness increased, with highest sharpness for the Bv89 kernel (2383.4 ± 787.1 ΔHU/mm). Blooming artifacts continuously decreased for reconstructions with the Bv40 (slice thickness, 0.2 mm; 52.8% ± 9.2%) to the Bv72 kernel (39.7% ± 9.1%). Subjective noise was perceived by both readers in agreement with the objective measurements. Considering delineation of coronary artery plaques and the adjacent vessel lumen, reconstructions with the Bv64 and Bv72 kernel (for both, median score of 5) were favored by the readers providing an excellent anatomic delineation of plaque characteristics and vessel lumen.

CONCLUSIONS: Ultra-high-resolution CCTA with PCD-CT is feasible and enables the visualization of calcified coronaries with an excellent image quality, high sharpness, and reduced blooming. Coronary plaque characterization and delineation of the adjacent vessel lumen are possible with an optimal quality using Bv64 kernel, a FOV of 200 × 200 mm 2 , and a matrix size of 512 × 512 pixels.

来源:Invest Radiol,查看原文。版权归原作者所有。