超声引导衰减参数与磁共振成像质子密度脂肪分数个体内比较:基于偏倚与精度统计
研究纳入309例同时接受UGAP和MRI-PDFF测量的患者,通过Bland-Altman分析进行个体内比较。结果显示偏倚为-0.0047,上下一致性界限分别为0.1160和-0.1255;百分比差异均值-1.1434%,上下一致性界限18.1723%和-20.4590%,计算期望一致性界限为18.5449%,283/309例(91.6%)的百分比差异在18.5449%以内。Bland-Altman分析显示UGAP与MRI-PDFF在临床可接受范围内可互换。
要点
- UGAP与MRI-PDFF个体内比较偏倚-0.0047,上下一致性界限0.1160和-0.1255。
- 百分比差异均值-1.1434%,上下一致性界限18.1723%和-20.4590%,期望一致性界限18.5449%。
- 283/309例(91.6%)百分比差异在18.5449%以内,两法在临床可接受范围内可互换。
核心结论
- Bland-Altman分析显示UGAP与MRI-PDFF在临床可接受范围内可互换。
研究设计与样本
- 纳入309例同时接受UGAP和MRI-PDFF测量的患者,将MRI-PDFF值转换为正态分布并通过线性回归统一单位后行Bland-Altman分析。
诊断标准与关键数值
- 偏倚-0.0047,上限一致性界限0.1160,下限一致性界限-0.1255。
- 百分比差异均值-1.1434%,上限一致性界限18.1723%,下限一致性界限-20.4590%。
- 计算期望一致性界限为18.5449%,283/309例(91.6%)百分比差异在18.5449%以内。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
Ultrasound-based techniques using the attenuation coefficient, including the ultrasound-guided attenuation parameter (UGAP), have been developed for the quantification of hepatic steatosis. The magnetic resonance imaging-based proton density fat fraction (MRI-PDFF) is considered to be more accurate than liver biopsy for liver fat quantification. The aim of this study was to perform intra-individual comparisons of UGAP and MRI-PDFF for determining hepatic steatosis grade. The study enrolled 309 patients who underwent UGAP and MRI-PDFF measurements. Bland-Altman analysis was conducted after transforming MRI-PDFF values to a normal distribution and converted to a common set of units using linear regression analysis for differing scales. The expected limits of agreement (LOA) was defined as the square root of the sum of the squares of UGAP and MRI-PDFF precision. A Bland-Altman plot revealed that the bias and upper and lower LOAs (ULOA and LLOA) were -0.0047, 0.1160 and -0.1255, respectively. The percentage difference indicated that the mean, ULOA and LLOA were -1.1434%, 18.1723% and -20.4590%, respectively. The calculated expected LOA was 18.5449%, and 283 of 309 patients (91.6%) had a percentage difference within 18.5449%. Bland-Altman analysis revealed that UGAP and MRI-PDFF were interchangeable within a clinically acceptable range.
引用
- Kumada, Ogawa, Goto, et al. Intra-individual Comparisons of the Ultrasound-Guided Attenuation Parameter and the Magnetic Resonance Imaging-Based Proton Density Fat Fraction Using Bias and Precision Statistics. Ultrasound Med Biol. 2022. PMID: 35613974. doi:10.1016/j.ultrasmedbio.2022.03.019
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。