文献速递乳腺乳腺X线

2016至2019年英国乳腺筛查项目中的辐射剂量

Radiation doses in the United Kingdom breast screening programmes 2016-2019.
PubMed 文献 · Br J Radiol · 2022-05-30

本研究汇总英国68,998名女性、411台X射线设备的乳腺摄影曝光参数,用英国标准方法估算平均腺体剂量(MGD)。结果显示数字乳腺摄影MGD较2010-12年升高11%,平均压迫乳腺厚度(CBT)增加(CC 4.8%、MLO 5.2%),MGD升高与CBT变化一致。

要点

  • 数字乳腺摄影平均MGD较2010-12年升高11%,MLO与CC视图均升高。
  • 平均CBT同期增加:CC视图4.8%,MLO视图5.2%。
  • MLO平均CBT为62.4±0.1 mm,超出诊断参考水平所用50至60 mm范围。

核心结论

  • 平均CBT随时间增加,导致数字乳腺摄影平均MGD升高;CBT亦随年龄变化。

研究设计与样本

  • 汇总英国411台X射线设备、68,998名女性的乳腺摄影曝光参数,用各设备局部输出量与半值层测量值按英国标准方法估算MGD。

诊断标准与关键数值

  • 数字乳腺摄影平均MGD较2010-12年升高11%,MLO与CC视图均升高。
  • 平均CBT较2010-12年增加:CC视图4.8%,MLO视图5.2%。
  • MLO平均CBT为62.4±0.1 mm,超出诊断参考水平所用50至60 mm范围。
  • CBT在50至60 mm范围时,MLO视图平均MGD为1.44±0.03 mGy。
  • CBT随年龄变化,峰值出现在51岁。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

英文原文(摘要)· 点击展开

OBJECTIVES: To record the radiation doses involved in UK breast screening and to identify any changes since previous publications related to technical factors and the population screened.

METHODS: Mammographic exposure factors for 68,998 women imaged using 411 X-ray sets spread across the UK were compiled. Local output and half value layer measurements for each X-ray set were used to estimate mean glandular dose (MGD) using the standard UK method.

RESULTS: Mean MGDs in digital mammography have increased by 11% since 2010-12 for both medio-lateral oblique (MLO) and cranio-caudal (CC) views. The mean compressed breast thickness (CBT) has increased (4.8% CC, 5.2% MLO) over the same period. The mean MLO CBT value of 62.4 ± 0.1 mm is outside the 50 to 60 mm range used for diagnostic reference levels. The increase in MGD is consistent with the CBT changes. The mean MGD in the 50 to 60 mm CBT range is 1.44 ± 0.03 mGy for MLO views. CBT varies with age and peaks at 51.

CONCLUSIONS: Mean CBT has increased with time, and this has increased mean MGDs for digital mammography. CBT also varies with age.

ADVANCES IN KNOWLEDGE: Updated average MGDs in the UK are provided. There is evidence that breast size is increasing in the UK and that mean CBT is affected by age-related changes in the breast.