物理上不可信信号作为前列腺弥散加权MR成像的定量质量评估指标
本研究通过识别明确代表数据伪影的物理上不可信信号(PIDS),对前列腺弥散加权MR图像进行定量评估,计算PIDS的百分比和分布,并比较使用与不使用直肠内线圈的mpMRI图像中PIDS的数量。结果显示直肠内线圈队列PIDS百分比为17.1±8.1%,无直肠内线圈队列为22.2±15.5%,两者无显著差异(p=0.14);直肠内线圈中中部PIDS(13.8±8.6%)低于尖部(19.9±11.1%)和底部(17.5±8.3%)(p=0.02)。结论认为PIDS图提供了空间分辨的定量质量评估,可指导放射科医生评估图像和改进DWI序列开发。
要点
- PIDS定义为扩散信号随b值增加而增加和/或ADC超过3.0 μm2/ms的体素。
- 直肠内线圈与无直肠内线圈队列整体PIDS百分比无显著差异(p=0.14)。
- 直肠内线圈中中部PIDS低于尖部和底部(p=0.02)。
核心结论
- PIDS图提供前列腺DWI的空间分辨定量质量评估;整体PIDS在直肠内线圈与无直肠内线圈队列间相似且各解剖区无显著差异,但许多患者特定前列腺区PIDS局灶性更高;PIDS评估可指导放射科医生评估图像及改进DWI序列开发。
研究设计与样本
- IRB批准的回顾性研究(2014年3月3日至2020年3月10日),纳入40例使用直肠内线圈(ERC)和40例不使用直肠内线圈(NERC)扫描的患者,共80例男性(58±8岁),使用本院MATLAB程序计算PIDS,由腹部影像专科培训且前列腺MRI阅片经验超过10年的放射科医生评估,本研究仅在前列腺中测试。
关键影像表现/征象
- PIDS包含扩散信号随b值增加而增加的体素和/或ADC超过3.0 μm2/ms的体素。
- 直肠内线圈队列PIDS体素百分比为17.1±8.1%,无直肠内线圈队列为22.2±15.5%。
- 直肠内线圈中中部PIDS为13.8±8.6%,低于尖部19.9±11.1%和底部17.5±8.3%(p=0.02)。
诊断标准与关键数值
- PIDS定义:任何体素满足(1)扩散信号随b值增加而增加和/或(2)表观扩散系数(ADC)超过3.0 μm2/ms(37°C纯水ADC,任何材料ADC更高在物理上不可信)。
- 无直肠内线圈与直肠内线圈PIDS无显著差异(p=0.14)。
- 直肠内线圈中尖部与底部PIDS百分比相似(p=0.30),无直肠内线圈中尖部与底部相似(p=0.86)。
临床应用与阅片要点
- PIDS图可提供前列腺DWI的空间分辨定量质量评估。
- PIDS评估可指导放射科医生评估图像及改进DWI序列开发。
局限性
- 本研究仅在前列腺中测试。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
PURPOSE: To provide a quantitative assessment of diffusion-weighted MR images of the prostate through identification of PIDS which clearly represents artifacts in the data. We calculated the percentage and distribution of PIDS in prostate DWI and compare the amount of PIDS between mpMRI images obtained with and without an endorectal coil.
METHODS: This IRB approved retrospective study (from 03/03/2014 to 03/10/2020), included 40 patients scanned with endorectal coil (ERC) and 40 without ER coil (NERC). PIDS contains any voxel where: (1) the diffusion signal increases despite an increase in b-value; and/or (2) apparent diffusion coefficient (ADC) is more than 3.0 μm2/ms (the ADC of pure water at 37 °C and it is physically implausible for any material to have a higher ADC). PIDS for transition zone (TZ) and peripheral zone (PZ) was calculated using an in-house MATLAB program. DWI images were quantitatively inspected for noise, motion, and distortion. T-test was used to compare the difference between PIDS levels in ERC versus NERC and ANOVA to compare the PIDS levels in the anatomic zones. The images were evaluated by a fellowship-trained radiologist in Abdominal Imaging with more than 10 years of experience in reading prostate MRI. This was tested only in prostate in this study.
RESULTS: 80 patients (58 ± 8 years old, 80 men) were evaluated. The percentage of voxels exhibiting PIDS was 17.1 ± 8.1% for the ERC cohort and 22.2 ± 15.5% for the NERC cohort. PIDS for NERC versus ERC were not significantly different (p = 0.14). The apex and base showed similar percentages of PIDS in ERC (p = 0.30) and NERC (p = 0.86). The mid (13.8 ± 8.6%) in ERC showed lower values (p = 0.02) of PIDS compared to apex (19.9 ± 11.1%) and base (17.5 ± 8.3%).
CONCLUSION: PIDS maps provide a spatially resolved quantitative quality assessment for prostate DWI. Average PIDS over the entire prostate were similar for the ERC and NERC cohorts, and did not differ significantly across prostate zones. However, for many of the patients, PIDS was focally much higher in specific prostate zones. PIDS assessment can guide Radiologist's evaluation of images and the development of improved DWI sequences.