乳腺X线筛查中自助加压与标准加压模式的技术参数及女性体验比较:一项单盲随机临床试验
在西班牙巴塞罗那Hospital del Mar开展的前瞻性随机对照研究纳入448名50至69岁无症状筛查女性,每名女性一侧乳房用标准加压(SC)、另一侧用患者辅助加压(PAC),共分析1792幅图像。PAC较SC获得更高压力(99.27 N vs 83.25 N,p<0.001),乳房厚度(56.11 mm vs 57.52 mm,p=0.015)和腺体剂量(1.34 mGy vs 1.37 mGy,p=0.018)更低;PAC不适评分略高(均值3.94 vs 3.69,p=0.042),但满意度调查中更多女性报告PAC引起更少不适,63.2%(289/448)女性偏好PAC。
要点
- 自助加压较标准加压获得更高压力(99.27 N vs 83.25 N,p<0.001)。
- 自助加压的乳房厚度和腺体剂量均低于标准加压(56.11 mm vs 57.52 mm;1.34 mGy vs 1.37 mGy)。
- 63.2%(289/448)女性偏好自助加压,更多女性报告其引起更少不适。
核心结论
- PAC在不影响其他技术成像参数的前提下获得更高压力,并改善女性筛查体验;作者认为两种模式间不适严重程度无临床显著差异。
研究设计与样本
- 前瞻性随机对照研究,在西班牙巴塞罗那Hospital del Mar开展,纳入448名50至69岁无症状筛查女性,筛查轮次为2017年12月至2019年12月,每名女性一侧乳房用SC、另一侧用PAC,按随机列表分配,共获取1792幅图像。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
英文原文(摘要)· 点击展开
OBJECTIVES: We compared the compression force, breast thickness, and glandular dose, as well as the severity of discomfort and women's experience between the patient-assisted compression (PAC) and standard compression (SC) modes.
MATERIALS AND METHODS: We conducted a prospective randomized controlled study at Hospital del Mar in Barcelona, Spain. We included 448 asymptomatic women aged 50 to 69 years old, attending their screening round from December 2017 to December 2019. Mammograms included the two bilateral views. In each woman, one breast was studied with SC and the other with PAC. The mode used in each breast was selected following a randomized list. Compression force, breast thickness, and average glandular dose were obtained for each of the 1792 images. We also recorded the degree of discomfort and women's experience, after mammogram acquisitions, using a predefined survey.
RESULTS: Higher compression forces were obtained with PAC than with SC (99.27 N vs 83.25 N, p < 0.001). Breast thickness mode (56.11 mm vs 57.52 mm, p = 0.015) and glandular dose (1.34 mGy vs 1.37 mGy, p = 0.018) were lower in PAC. The discomfort score was slightly higher with PAC (mean 3.94 vs 3.69, p = 0.042), but in the satisfaction survey, more women reported that PAC caused less discomfort. Additionally, 63.2% of women (289/448) preferred PAC.
CONCLUSION: PAC achieved higher compression forces without impairing the other technical imaging parameters and enhanced women's experience of screening mammography. We believe there were no clinically significant differences in the severity of discomfort between the two modes.
KEY POINTS: • Self-compression allows higher compression forces than the standard compression mode. • Self-compression does not affect technical imaging parameters. • Self-compression improved women's experience of screening mammography when standard compression was used on one breast and self-compression on the other.
引用
- Arenas, Alcantara, Posso, et al. Comparison of technical parameters and women's experience between self-compression and standard compression modes in mammography screening: a single-blind randomized clinical trial. Eur Radiol. 2022. PMID: 35536390. doi:10.1007/s00330-022-08835-y